Do Not Touch Test Leads while test is in Progress!
71-023 Rev C 5/8/2007
15) OUTPUT CONTROL
This control adjusts the output voltage of the tester. Clockwise (CW)
rotation increases output and counterclockwise (CCW) rotation
decreased output. Full CCW is the MIN (minimum) point output and
full CW is MAX (maximum) or 100% of the testers rated output.
Note:
A “Zero Start Interlock” is connected to the output control for
use during Surge and HiPot testing. The user must return the output
control to MIN each time before pressing the Test button (6). In other
words, if the Test button is pressed while the Output control is above
zero the instruments output is disabled until the Output control is
rotated to MIN.
16) FOOTSWITCH CONNECTOR (REAR PANEL)
A footswitch may be connected to this socket which is in parallel to the
Test button. The footswitch will operate the tester in a manner
identical to the Test button, freeing the user’s hands from having to
operate the switch.
17) SURGE/HIPOT SELECTOR. This function, like Function selector
(12), is a relay that selects the type of test, Surge or HiPot, to be performed.
The name of the test chosen is displayed on the lower portion of the CRT
for reference.
For testing windings and stators, it should always agree with the
position of the Function Selector.
Three Phase Selector Switch
This is an optional high-voltage rotary switch that selects which test
lead is hot, open, or ground. If this switch is present on the tester, refer
to Appendix D: Three Phase Operation for a description and
instructions to use this switch for surge testing.
18) LEADS ENERGIZED INDICATOR LIGHT
This indicator light will illuminate when voltage is applied during a
test.
19) TEST LEADS (LOWER SECTION)
Test leads, one red and three black or ground leads, are provided for
contact to the windings. Test leads are insulated to 60kV.
Summary of Contents for D30R
Page 2: ...D30R Users Manual 2 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 4: ...D30R Users Manual 4 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 10: ...D30R Users Manual 10 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 16: ...D30R Users Manual 16 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 24: ...D30R Users Manual 24 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 32: ...D30R Users Manual 32 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 34: ...D30R Users Manual 34 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 40: ...D30R Users Manual 40 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 46: ...D30R Users Manual 46 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 62: ...D30R Users Manual 62 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 68: ...D30R Users Manual 68 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 70: ...D30R Users Manual 70 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 80: ...D30R Users Manual 80 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 82: ...D30R Users Manual 82 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 106: ...D30R Users Manual 106 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 108: ...D30R Users Manual 108 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 118: ...D30R Users Manual 118 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 120: ...D30R Users Manual 120 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 124: ...D30R Users Manual 124 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 134: ...D30R Users Manual 134 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 136: ...D30R Users Manual 136 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...
Page 140: ...D30R Users Manual 140 Do Not Touch Test Leads while test is in Progress 71 023 Rev C 5 8 2007...