Everest Ca-Zoom HD
Page 12-2
3068-7-0001 Rev. d Version 1.0
12.2
Maintenance work and functional test
NOTICE
Regular cleaning, care and maintenance within the intervals
described below will ensure efficient operational readiness.
All maintenance work must be performed by trained and authorized staff in
compliance with the instruction manual, taking all safety instructions into
consideration.
12.2.1
Inspections before each use
Cables: Visually inspect the cables for integrity (connector outlet and
outer sheath).
Assemblies, screws and clamps: Make sure that the mounted parts and
screws are not loose or damaged.
12.2.2
Apply compressed air to the device
WARNING
Danger from bursting camera head.
Camera head explodes when over pressurized.
Wear protective goggles when repressurizing camera
head.
Only use supplier CO
2
injector to pressurize camera
head. Do not pressurize directly from any other source.
Only use pressurization cartridges approved by supplier.
Do not puncture CO
2
cartridge or heat above
49 °C/120 °F.
Compressed air or CO
2
must be applied to the camera head for underwater use.
This protects the high-quality electronics in the event of leakage. The camera
head either indicates the loss of tightness by periodically switching off the lights
or by an alert on the control software (newer models).
Summary of Contents for Waygate Everest Ca-Zoom HD
Page 1: ...Everest Ca Zoom HD Operating Instructions...
Page 2: ...Everest Ca Zoom HD Page ii 3068 7 0001 Rev d Version 1 0...
Page 4: ......
Page 29: ...Everest Ca Zoom HD 3068 7 0001 Rev d Version 1 0 Page 3 3 Figure 10 Typical Ca Zoom HD System...
Page 30: ......
Page 32: ......
Page 44: ......
Page 46: ......
Page 60: ......
Page 64: ......
Page 66: ......
Page 74: ......
Page 76: ......
Page 80: ......