Appendix A. Outline and Installation Drawings
24
oxy.IQ User’s Manual
Figure 9: Outline and installation (ref. dwg. 712-1840)
Figure 10: Cable, standard (ref. dwg. 704-1317, SH 1 of 2)
Connector
1.58
[40.3]
2.05
[52.1]
B
C
2.75
[69.8]
4.10
[104.2]
1.33
[33.9]
A
0.32
[8.25]
1.03
[26.27]
0.53
[13.57]
1.00
[25.40]
2X
1/8-27NPT-2B
.27 [6.9]
2X
8-32 UNC-2B
.51 [12.9] max
Ø1.75
[44.3]
Notes:
1. Weight: 1.35 lb [612.3 g]
2. Color: metallic
3. Dimensions: inches [mm]
A
Ø .03 [.76]
Ø .03 [.76]
C
B
A
C
B
.25±.03
XX
1.16
1.00±.25
A
A
1
2
3
Blue
Brown
4
1
2
4
3
.49
Indicator arrow
1.00
Section A-A
Scale 2 : 1
3
Cable length
5
Heat shrink with adhesive backing
XX-cable length
Part number
extension
2 meters -0/.1 (6.56ft -0/+.33)
02
10 meters -0/.5 (32.81ft -0/+1.58)
10
End “A”
End “B”
Notes:
1. Interpet drawing in accordance with ASME Y14.5(M)-2009
2. Finished components to be ROHS compliant
3. Part to be clean of dirt/debris
4. Performance specification
Rated current 2.0 A
Rated voltage 24 VDC
Operating temperature range -25 C TO 85 C
Protection class meets NEMA type 1,3,4,6P AND IEC IP67
Retention 6.75 lbs. min. perpendicular & normal
5. Cable assembly shall be tested 100% for continuity and shorts.
Hirose solder & back shell tightening fixture shown on sheet 2.
6. Inspection requirements:
AllCTQs (1-11) identified & notes are to be reported on the FAI & meet AQL .65 level II per lot
Specific CTQs identified below are to have a CpK no less than 1.33:
1,2,4,10
Specific CTQs identified below are to have a Cp no less than 1.33:
@LSL - none @USL - none
Attribute data to meet 4 SIGMA (0.63% DPMO)identified below:
None
Cp & CpK, DPMO
Summary of Contents for Panametrics oxy.IQ
Page 1: ...panametrics com BH024C11 EN G oxy IQ Panametrics Oxygen Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter 1 Features and Capabilities 4 oxy IQ User s Manual no content intended for this page...
Page 26: ...Chapter 5 The Service Menu 18 oxy IQ User s Manual no content intended for this page...
Page 30: ...Chapter 6 Specifications 22 oxy IQ User s Manual no content intended for this page...
Page 42: ...Appendix C Order String 34 oxy IQ User s Manual no content intended for this page...
Page 50: ...Appendix E Cell Models 42 oxy IQ User s Manual no content intended for this page...
Page 54: ...Warranty 48 oxy IQ User s Manual no content intended for this page...
Page 55: ......