Chapter 4. User Programming
16
oxy.IQ User’s Manual
4.4.4 Error Output
To select the desired output value that will be sent to the analog output device upon an error, complete the following
steps:
1.
Press the
Enter
key to enter the
Main Menu
.
2.
Press the
key twice and then press the
Enter
key to enter the
Output
menu.
3.
Press the
key three times and then press the
Enter
key to enter the
Error Output
menu.
4.
Use the
and
keys to select the desired option and then press the
Enter
key to activate that error
output. A check mark will appear next to the selected option to indicate that it is activated. The following options
are available, and you may activate only one option at a time.
Note:
In order to implement the Namur output, two settings must be changed:
1.
Set the output error type to low O
2
(Main Menu –> Output –> Error type –> Low O
2
–> Enter
).
2.
Activate the NAMUR over the Error output menu and select NAMUR (Main Menu –> Output –> Error output –>
NAMUR –> Enter
).
Note:
Only the first four options are displayed on the screen upon entering this menu. When you scroll down to the
fourth option (NAMUR), a down arrow to the right of this option indicates that an additional screen of options
is available.
• None (no error output is generated)
• Low (an output below 4 mA is generated)
• High (an output above 20 mA is generated)
• Value (an error output at a programmable fixed value is generated)
• NAMUR (listed on second screen of options)
Note:
Pressing the
Enter
key on a different error output will automatically deselect any previously selected
output.
3.
Press the
Cancel
key three times to return to measurement mode.
Summary of Contents for Panametrics oxy.IQ
Page 1: ...panametrics com BH024C11 EN G oxy IQ Panametrics Oxygen Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter 1 Features and Capabilities 4 oxy IQ User s Manual no content intended for this page...
Page 26: ...Chapter 5 The Service Menu 18 oxy IQ User s Manual no content intended for this page...
Page 30: ...Chapter 6 Specifications 22 oxy IQ User s Manual no content intended for this page...
Page 42: ...Appendix C Order String 34 oxy IQ User s Manual no content intended for this page...
Page 50: ...Appendix E Cell Models 42 oxy IQ User s Manual no content intended for this page...
Page 54: ...Warranty 48 oxy IQ User s Manual no content intended for this page...
Page 55: ......