oxy.IQ User’s Manual
21
Chapter 6. Specifications
6.3.15 European Compliance
See the EU declaration of conformity at the back of this manual.
6.4 Storage
•
The storage environment should be cool and dry.
•
The ideal temperature range is 0°C to 25°C (32°F to 77°F).
•
The fuel cell can be intermittently stored at a temperature of up to 55°C (131°F). (e.g., during transportation).
6.5 Product label
A typical product label is shown in figure 8 below.
Figure 8: oxy.IQ label - IS package option
II 1 G
Panametrics
Free Zone East, Shannon,
Co. Clare, V14 V992, Ireland
IS CLI, Div 1, GPS ABCD
CLI. Div 2, GPS ABCD
Class 1, Zone 0. AEx/Ex ia IIC
T4 Ga
Tamb -20°C to +60°C
Install per control drawing
752-347
Ex ia IIC T4 Ga
Tamb -20°C to +60°C
Ui = 28V, Ii = 150mA, Pi = 1.05W
Ci = 0, Li = 0
FM14ATEX0032X
IECExFMG14.0016X
Model: OXYIQ-XXX-XX
SN: XXXX
Mfg date: Month year
0598
EN50104
MSIP-REI-PK5-oxyIQ
oxyIQ
Summary of Contents for Panametrics oxy.IQ
Page 1: ...panametrics com BH024C11 EN G oxy IQ Panametrics Oxygen Transmitter User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter 1 Features and Capabilities 4 oxy IQ User s Manual no content intended for this page...
Page 26: ...Chapter 5 The Service Menu 18 oxy IQ User s Manual no content intended for this page...
Page 30: ...Chapter 6 Specifications 22 oxy IQ User s Manual no content intended for this page...
Page 42: ...Appendix C Order String 34 oxy IQ User s Manual no content intended for this page...
Page 50: ...Appendix E Cell Models 42 oxy IQ User s Manual no content intended for this page...
Page 54: ...Warranty 48 oxy IQ User s Manual no content intended for this page...
Page 55: ......