background image

Chapter 3: Service and maintenance

3.1  The service menus

IMPORTANT

The service menus are accessible only by using the 

factory-level passcode.

After entering the factory 

level passcode, use the 

arrow keys to scroll to 

service and press enter.

To check the version levels 

of your HygroPro firmware, 

select versions and press 

enter. The information for 

your unit is displayed.

The following HygroPro service menu options are also 

available:
•  Upgrade - used to install an updated firmware version
•  Default analyzer - used to reset all settings to the  

factory defaults

Select the desired service 

menu option and press 

enter. Then, follow the on-

screen instructions.

After completing the 

above step, respond to the 

ARE YOU SURE? question. 

Be sure to respond yes, if 

you wish to save the new 

information.

3.2  Moisture probe error conditions

IMPORTANT

All moisture probes require periodic cleaning to maintain 

optimum accuracy. Consult the Panametrics service 

center for the recommended probe cleaning interval for 

your application.

If there is a problem with the moisture probe during 

operation, the HygroPro is programmed to indicate the  

error condition via its analog output signal. To indicate a 

probe error condition, the analog output signal is forced  

to the following values:
•  ≥22 mA to indicate a short circuit in the probe
•  ≤3.5 mA to indicate an open circuit in the probe
If you have one of the above error conditions, follow the 

probe cleaning instructions in the next section.

16

Summary of Contents for Panametrics HygroPro

Page 1: ...HygroPro User s manual 916 099 Rev D...

Page 2: ...ii...

Page 3: ...HygroPro Moisture transmitter User s manual 916 099 Rev D November 2012 iii...

Page 4: ...no content intended for this page proceed to next page iv...

Page 5: ...1 2 2 3 Selecting measurement parameters 11 2 2 4 Setting up the display 12 2 2 5 Setting up the analog output 12 2 2 6 Entering the node ID 13 2 2 7 Setting up the RS485 output 13 2 3 Advanced setup...

Page 6: ...the cleaned probe 18 Chapter 4 Specifications 19 4 1 General 19 4 2 Electrical 19 4 3 Mechanical 19 4 4 Moisture sensor 19 4 5 Built in temperature sensor 20 4 6 Built in pressure sensor 20 4 7 Certi...

Page 7: ...t the following link https www bakerhughesds com sales terms conditions Typographical conventions Note These paragraphs provide information that provides a deeper understanding of the situation but is...

Page 8: ...ke sure that unauthorized personnel cannot gain access to the operation of the equipment Environmental compliance RoHS The HygroPro moisture transmitter fully complies with RoHS regulations Directive...

Page 9: ...x button keypad and it is housed in an IP67 Type 4X enclosure The HygroPro transmitter uses an aluminum oxide moisture sensor and it includes a temperature thermistor and a pressure transducer on a co...

Page 10: ...he sample flow Before constructing a sample system consult a Panametrics application engineer and observe the following guidelines The sample system should be very simple and should contain as few com...

Page 11: ...age during operation 2 Using the integral 3 4 16 straight male thread screw the probe end of the transmitter into the process or sample system fitting Make sure not to damage the threads Note A 3 4 16...

Page 12: ...receive and display the 4 20 mA measurement output from the HygroPro proceed to next page an external power supply to provide power to the HygroPro plus a personal computer PC running Panametrics Pan...

Page 13: ...ground 1 Connect an RS485 RS232 converter customer supplied to an available serial port on the PC 2 Push the female connector end of the factory supplied cable into the mating male connector on the t...

Page 14: ...da for the following hazardous classified locations and protection methods Intrinsically safe class I II III division 1 group ABCDEFG T4 T Ambient 40 C to 60 C Enclosure IP 67 3 Galvanically isolated...

Page 15: ...C22 2 No 61010 1 Degrees of Protection Provided by Enclosures IP Code ANSI IEC 60529 Degrees of Protection Provided by Enclosures IP Code CSA C22 2 No 60529 1 4 2b Compliance requirements The HygroPr...

Page 16: ...e end of the HygroPro ATEX certificate number Bas06ATEX0019X indicates that special conditions are required for safe operation in Europe These conditions are 1 The equipment must be protected against...

Page 17: ...plied to the unit The transmitter requires up to 60 seconds to initialize and begin normal operation The unit will meet its specified accuracy within 3 minutes Figure 7 below shows a close up view of...

Page 18: ...st temperature Node ID enter value Baud rate Parity Stop bits Data Bits Save Table RS485 Service Zero mV Zero PSIg Span mV Span PSIg Temp C 6 CS 6 Temp C 1 CS 1 Row 16 Units 16 Row 1 Units 1 Figure 8...

Page 19: ...level passcode must be entered 2 2 3 Selecting measurement parameters To select the measurement parameters complete the following steps To select the number of measurements to be displayed on each scr...

Page 20: ...he above screen highlight that choice and press enter Use the arrow keys to change the contrast value and press enter Then press escape twice to return to the main menu 2 2 5 Setting up the analog out...

Page 21: ...twork when used with PanaView interface software To enter your nNode ID proceed as follows Enter your passcode to access the setup program Then use the arrow keys to select the Node ID menu Press ente...

Page 22: ...o enter the desired pressure value and press enter The screen will be updated to show the new pressure value To set the temperature display enter your password and use the arrow keys to select const t...

Page 23: ...to select lock keypad on the initial screen and press enter twice The keypad is now locked To unlock the keypad just re enter the user program as described on page 11 No further programming steps are...

Page 24: ...creen instructions After completing the above step respond to the ARE YOU SURE question Be sure to respond yes if you wish to save the new information 3 2 Moisture probe error conditions IMPORTANT All...

Page 25: ...al depends on the specific application To accomplish this either return the RTE to Panametrics for recalibration or install a new RTE The HygroPro electronics will automatically read and store the cal...

Page 26: ...n set at 50 C 2 C 122 F 3 6 F for 24 hours 3 3 6 Installing the probe in the transmitter To install a new or cleaned probe in the transmitter complete the following steps 1 Reconnect the probe cable t...

Page 27: ...iven a 24 VDC power supply Max load resistance 24 x 33 33 300 500 Cable 6 ft 2 m standard consult Panametrics for custom lengths Input parameters for loop powered intrinsic safety Ui 28V Pi 0 653W Li...

Page 28: ...lable ranges 30 to 300 psig 3 to 21 bar 50 to 500 psig 4 to 35 bar 100 to 1000 psig 7 to 69 bar 300 to 3000 psig 21 to 207 bar 500 to 5000 psig 35 to 345 bar Note Psig ranges are based on a constant p...

Page 29: ...s arising from course of dealing or usage or trade Return policy If a Panametrics instrument malfunctions within the warranty period the following procedure must be completed 1 Notify Panametrics givi...

Page 30: ...uction targets across critical industries including Oil Gas Energy Healthcare Water and Wastewater Chemical Processing Food Beverage and many others Join the conversation and follow us on LinkedIn lin...

Reviews: