Enter the
zero
and
span
values for the range,
using the same procedure
on the previous page.
In the output menu, select trim and press
enter
. The
following screen appears:
Enter your values for
the base trim, span trim
and %, using the same
procedures as on the
previous page.
When you select the
mode
option at the above prompt,
choose either test to verify the output or normal for normal
operation.
2.2.6 Entering the node ID
The
node ID
is a unique network identifier that enables
the
HygroPro
to be connected to a multi-drop network
when used with PanaView™ interface software. To enter
your nNode ID, proceed as follows:
Enter your passcode to
access the setup program.
Then use the arrow keys to
select the Node ID menu.
Press
enter
.
Use the arrow keys to enter
the desired
node ID
value
and press
enter
.
2.2.7 Setting up the RS485 output
To set up the RS485 output, complete the following steps:
Enter your passcode to
access the setup program.
Then use the arrow keys
to select the RS485 menu.
Press
enter
.
To enter the
baud rate
,
press
enter
.
Use the up and down keys
to select the desired value
and press
enter
.
Repeat the above procedure to enter the following RS485
parameters:
• Parity
• Stop bits
• Data bits
You have now completed the initial setup process.