Aurora Rack Mount Manual Adendum

5

 

Figure 4: Wiring Diagram for DB-15 I/O connector

Use shielded 18-22 AWG (0.82-0.33 mm2) twisted pair wire for all I/O connections.  To ensure compliance to EN 61010 

requirements, a shielded cable is required for these connections and the shield must be terminated to the Ground  

( ) terminal on the back panel. 
For operation on RS-485, 2-wire, half-duplex bus, attach the RS-485 (+) to (+) and the RS-485 (-) to (-).  If available, 

make a third connection to the RS-485 network common to RTN. Connect the shield wire from the cable to SHD 

terminal.  For operation on RS-232, connect RS-232 (TXD) to (+), RS-232 (RXD) to (-), and RS-232 (GND) to RTN. 

Connect the cable shield wire to SHD terminal.

Note:

The default configuration of the two serial ports as shipped is shown below:

Table 4: RS-232/RS-485 wiring table

Aurora Serial Terminal

Color code

(Port 1 Pin) (Port 2 Pin)

White

TX

+ (10) (14)

Red

RX

- (11) (15)

Green

Return

RTN (12) (16)

Black

Shield

SHD (13) (13)

Table 5:  Default Serial port configuration

BAUD Rate:

115,200

Parity:

Even

ID Note:

1 for Port 1, 2 for Port 2

Summary of Contents for Panametrics Aurora RM

Page 1: ...H2O User manual P N BH003C11 rev C for operating the Aurora RM Please refer to the Aurora H2O User Manual for detailed instructions on operation general programming maintenance and troubleshooting in...

Page 2: ......

Page 3: ...material contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries All third party product and company names are trademarks of their respective h...

Page 4: ...ii no content intended for this page...

Page 5: ...ervices including Training Product Repairs Service Agreements and more Please visit https www bakerhughes com panametrics panametrics services for more details CAUTION This symbol indicates a risk of...

Page 6: ...at you bought has required the extraction and use of natural resources for its production It may contain hazardous substances that could impact health and the environment In order to avoid the dissemi...

Page 7: ...Preface v no content intended for this page...

Page 8: ...mounting in a standard 19 rack for quick deployment and use The basic sample conditioning system allows for quick and easy gas connections on the rear panel flow metering and adjustment on the front...

Page 9: ...r horizontal flow cell is purged Dry down from ambient 5000 PPMv to 10 PPMv in seconds which enables very fast recovery from dehydration system upsets Direct readout in PPMv dew frost point C F densit...

Page 10: ...Aurora Rack Mount Manual Adendum 3 2 Mounting the Aurora RM Rack Mount Analyzer Figure 2 Aurora Drawing for Rack Mounting P N 165M1371 rev A...

Page 11: ...r user manuals for proper installation 3 Wiring the Aurora Rack Mount Configuration All the electrical connections input and output to the Aurora RM are made via the back panel as shown in Figure 3 be...

Page 12: ...85 to If available make a third connection to the RS 485 network common to RTN Connect the shield wire from the cable to SHD terminal For operation on RS 232 connect RS 232 TXD to RS 232 RXD to and RS...

Page 13: ...dous concentrations of flammable toxic or inert gases For safe and effective operation both the gas inlet and vent connections should be leak checked as ambient air leaks into the sample flow can lead...

Page 14: ...sity of the laser beam or the accuracy of the moisture measurement When the Weak Signal error appears on the LCD display the laser absorption cell may need some maintenance Please contact Panametrics...

Page 15: ...ewdriver and pry the fuse holder loose until your fingers can be used as shown in Figure 9 Gently pull on the fuse holder until it stops Swivel the fuse holder carefully as shown below and replace fus...

Page 16: ......

Page 17: ...eir respective holders BH005C11 EN D 09 23 Scan here for Panametrics support and service information Customer Support Centers U S A The Boston Center 1100 Technology Park Drive Billerica MA 01821 U S...

Reviews: