5. Maintenance and troubleshooting
The Aurora RM has minimal routine or preventative maintenance requirements if used per instructions provided in
this user manual addendum. Since the laser absorption cell is mounted horizontally, any accidental liquid carryover
into it is likely to settle at the bottom of the cell without affecting the intensity of the laser beam or the accuracy of the
moisture measurement. When the "Weak Signal" error appears on the LCD display, the laser absorption cell may need
some maintenance. Please contact Panametrics Technical Support and do not attempt to access and/or clean the
absorption cell.
WARNING!
Leaks of combustible and/or toxic gases could occur over the life of the analyzer.
Standard industry practices to monitor and detect hazardous gas leaks and take
appropriate safety actions must be followed.
CAUTION!
Please be sure to power down the instrument before opening the top cover of the
rack-mount enclosure to perform any service/maintenance.