4.5.4 Editing pressure calibration
Pressure input in this section is used only for equivalent dew point calculations. Equivalent dew point is the dew point of the
process gas at the process pressure. Input a “constant” value if the line pressure is at a normal pressure, or use an external
pressure transmitter to input a “live” pressure input into the
Aurora H2O
analyzer.
To edit the pressure
calibration, from the line
pressure source menu,
select live and press enter.
The following screen
appears.
To edit the calibration
data, use the arrow keys to
select cal data and press
enter. The following screen
appears.
To select the calibration
point, use the up and
down arrow keys to
highlight select cal point
and press enter. The
following screen appears.
Use the left and right
arrow keys to select zero
or span and press enter.
The screen returns to the
previous menu.
To edit the pressure value,
use the up and down
arrow keys to select edit
pressure value and press
enter. The following screen
appears.
Use the left and right
arrow keys to select each
digit to be changed. Use
the up and down arrow
keys to change the value.
When finished, press enter.
The screen returns to the
previous menu.
To edit the input value, use
the up and down arrow
keys to select edit input
value and press enter. The
following screen appears.
Use the left and right
arrow keys to select each
digit to be changed. Use
the up and down arrow
keys to change the value.
When finished, press enter.
The screen returns to the
previous menu.
44