1.4.12 User interface
Programmable “through-the-glass” via magnetic stylus
1.4.13 Weight and dimensions
Net Weight
45 kg (100 lb)
Overall dimensions
34”H x 18”W x 14”D (87 cm x 46 cm x 36 cm)
[see drawings]
1.4.14 Hazardous area certification
USA/Canada
Explosion-proof for class I,
division 1, groups B, C&D
Dust-Ignitionproof for class II/III,
division 1, groups E, F&G
(Canada only)
EU and elsewhere
ATEX and IECEx: Ex d e IIB+H2 T6 Gb IP66;
Tamb = -20°C to +65°C. Flameproof with
increased safety compartment
Evaluated against standards EN 60079-0:2012,
EN 60079-1:2007, EN 60079-7:2007, IEC 60079-0:2011,
IEC 60079-1:2007, IEC 60079-7:2006.
1.4.15 European certification
CE
Complies with EMC directive 2014/30/EU low voltage directive
2014/35/EU and pressure directive 2014/68/EU for DN<25
1.4.16 Software
Displays all key parameters. provides time-base graphing,
data logging ability, export data as ASCII text. Software has
lockout/passcode.
1.4.17 Sampling system
Integrated sampling system to filter out physical
contaminants, remove glycol carry over, control
temperature condition, regulate pressure and flow rate.
Optional thermostatically controlled heater.
1.4.18 Wetted components
316/316L stainless steel for tubing and fittings. Other wetted
components such as the optical window and mirror are
proprietary materials and are compatible with natural
gas and typical contaminants found in natural gas. Other
sample system components wetted parts include PTFE,
PFA, Inconel, Hastelloy, PVDF, glass, Viton.
9
Summary of Contents for Panametrics Aurora H2O
Page 2: ...ii...
Page 4: ...no content intended for this page proceed to next page iv...
Page 20: ...no content intended for this page 10...
Page 32: ...no content intended for this page 22...
Page 71: ...Figure 74 Exporting a simplified image Figure 75 Selecting enhanced metafile 61...