Copyright 2022 Baker Hughes Company.
English–PTC Temperature Calibrators | 41
Observe the instructions in the safety data sheet of the calibration liquid.
11.3.4.1 Distilled Water
1. Remove the sensor cage from the tank.
2. Remove the magnetic stirrer with the aid of the magnetic lifter.
3. Thoroughly dry the micro bath, sensor cage, magnetic stirrer and the draining syringe.
11.3.4.2 Silicone Oil
1. Remove the sensor cage from the micro bath.
2. Remove the magnetic stirrer with the aid of the magnetic lifter.
3. Clean the micro bath, sensor cage, magnetic stirrer with water to which a generous amount
of detergent has been added.
4. Remove the soapy water as much as possible using the draining syringe.
5. Thoroughly dry the micro bath, sensor cage, magnetic stirrer and the draining syringe.
11.4 Return Goods/Material Procedure
If the unit requires calibration or is unserviceable, return it to the nearest Druck Service Centre
https://qrco.de/bcPHml
.
Contact the Service Department to obtain a Return Goods/Material Authorization (RGA or RMA).
Provide the following information for a RGA or RMA:
•
Product (e.g. Dry
PTC
700).
•
Serial number.
•
Details of defect/work to be undertaken.
•
Include any error code(s).
•
Operating conditions.
11.4.1 Safety Precautions
You must inform Druck if the product has been in contact with any hazardous or toxic substance.
The relevant COSHH or in the USA, MSDS, references and precautions to be taken when
handling.
12. Decommissioning and Disposal
12.1 Before Decommissioning
Prior to decommissioning, ensure that:
1. The measurement set up is switched off and is in a safe de-energized state.
2. The calibrator and its accessories have completely cooled down. See Section
12.2 Decommissioning
1. Remove all connected sensors and devices.
2. Switch off the calibrator and disconnect the mains plug.
3. Empty the micro bath of residual calibration liquid. See Section
INFORMATION
Service by unauthorized sources will affect the warranty and
may not guarantee further performance.