30 | PTC Temperature Calibrators–English
3. Confirm the selection.
Waiting Time
The waiting time is a manually entered time delay to allow the DUT to stabilize before data
acquisition.
Note:
The waiting time is only used if user input was selected for data acquisition.
Gradient
The gradient is used when the internal measuring instrument or TTScan is selected for data
acquisition. If the gradient falls below the set value, the DUT is stable.
Customer Specific Tolerance Value
Customer-specific tolerance values are used if the DUT has not been linked to a standard
tolerance class or if a standard signal (current or voltage) is used.
1. Tap “Tolerance class”.
2. Select customer-specific and confirm the selection.
3. Tap “Tolerance value”.
Tolerance value = Percent Value [% of reading] + Constant Value [constant]
4. Confirm the input.
6.2.5 Test Points
1. In the “Configure Test Task” window, see Figure
20, tap the “Test Points” icon.
Figure
24: Configure Test Points Window
6.2.5.1 Create Test Points
1. Tap the
icon.
2. Enter temperature and dwell time and confirm the respective input.
6.2.5.2 Edit Test Points
1. Tap the
icon.
INFORMATION
If the automatic data acquisition (Switch Test) is selected
and the DUT type is set to switch, the test points “Starting Temperature”, “Test
Area Beginning” and “Test Area End” are specified. “Test Point Calculation” is
not available.