Copyright 2022 Baker Hughes Company.
English–PTC Temperature Calibrators | 31
2. Change temperature and dwell time and confirm the new values.
6.2.5.3 Test Point Calculation
Use Test Point Calculation to automatically calculate the intervening test points between a
specified first and last test point.
Note:
When a parameter is changed, the other dependent parameters are automatically
recalculated.
1. Tap the
icon.
Figure
25: Calculate Test Point Window
Step, One-Sided
The one-sided step
calculates the intervening test points in succession from the first test
point to the last test point.
Step, Two-Sided
The two-sided step
calculates the intervening test points in succession from the first test
point to the last test point, then back to the first test point.
Define First and Last Test Points
1. Tap on the fields, enter the new values and confirm the input.
2. The intervening test points are automatically recalculated.
3. The screen returns to the previous window.
Cycles
Determines the number of cycles (repetitions) of a whole test sequence.
1. Tap the “Cycles”
icon.
2. Enter the new value and confirm the input.
3. The screen returns to the previous window.
Set Interval
For fixed intervals:
1. Tap the “Interval” field.
2. Enter the desired step width and confirm the input.
3. The number of steps are automatically calculated.
Note:
If necessary, the last test point is also adjusted to match.