12 | PTC Temperature Calibrators–English
4.3.1.1 Installation
The adaptor sleeve is inserted into the dry block using the insert exchange tool, see Figure
item
3.
4.3.1.2 External Reference Sensor (Optional Accessory)
Align the adapter sleeve so that the hole for the external reference sensor is located at 12 o’clock.
4.3.1.3 Removal and Cleaning
1. Allow the calibrator cool before the adaptor sleeve is removed.
2. Pull the adaptor sleeve out of the dry block with the help of the insert exchange tool, see
4 item
3.
3. Clean the adaptor sleeve and the dry block to prevent inserts from getting stuck in the dry
block.
4.3.2 Air Shield Insert
The air shield insert provides optimum radial and axial temperature distribution. The borehole
divider provides a flexible and cost-effective adaptation to various calibration tasks. For air shield
insert compatibility, see Section
8.
Figure
5: Air Shield Insert
4.3.2.1 Installation
The air shield is inserted into the dry block using the insert exchange tool, see Figure
5 item
2.
4.3.2.2 External Reference Sensor (Optional Accessory)
Align the air shield so that the hole for the external reference sensor is located at 12 o’clock.
4.3.2.3 Removal and Cleaning
1. Allow the calibrator cool before the air shield is removed.
2. Pull the air shield out of the dry block with the help of the insert exchange tool, see Figure
item
2.
3. Clean the air shield and the dry block to prevent inserts from getting stuck in the dry block.
INFORMATION
Use only Druck adaptor sleeves.
If in doubt, contact Druck.
1 Air shield insert.
2 Insert exchange tool.