4. The screen returns to the previous window.
Defining the Number of Steps
Note:
The total number of resulting test points is always one higher than the entered number of
steps.
1. Tap the field “Number of Steps” to change the number of steps.
2. Enter the new value and confirm the input.
3. The number of steps is applied and the interval is automatically recalculated.
Set Dwell Time
1. Tap the field “Dwell Time”.
2. Enter the new value and confirm the input.
3. The screen returns to the previous window.
Set Gradient
1. Tap the field “Gradient”.
2. Enter the new value and confirm the input.
3. The screen returns to the previous window.
Behavior at Test End
Set the behavior of the calibrator at the end of the test.
1. Tap the field “Behavior at Test End”.
2. The “Select Behavior at Test End” window appears.
Figure
26: Behavior at Test End Window
3. Select the desired behavior and confirm the input.
4. The screen returns to the previous window.
Safety Temperature Value
Note:
The “Safety Temperature Value” field is only visible if “Safety Temperature Value” has
been selected as the behavior at the end of the test. See “Behavior at Test End” on page
1. Tap the “Safety Temperature Value”
icon.
2. Enter the new value and confirm the input.
3. The screen returns to the previous window.