Copyright 2010 Baker Hughes Company.
30 | TERPS Instruction Manual–English
Appendix
A.
Fault Finding
!021 Bad
Checksum
Error occurred in memory check sum.
Power cycle sensor, if fault remains return
to Druck for repair or replacement.
!022 Bad
Message
Erroneous character detected in the
middle of a message.
Check and re-send message.
!023 Bad Cal
Pres
Divide by zero error occurred in
calibration calculation.
Check data and repeat calibration.
Table
A
-
1: Fault Finding
Error Message Probable cause
Action