Copyright 2010 Baker Hughes Company.
18 | TERPS Instruction Manual–English
Reply:
In direct mode:
<Pressure reading><Units><CR>
Result:
A pressure reading and the units of pressure measurement is
transmitted every <Interval> seconds in the format set in the set-up
commands. The interval input is non-volatile.
Query:
<SPACE>A,?<CR>
Reply:
<Transmission interval set>,<Y if Units displayed, N if units not
displayed><CR>
Query:
<SPACE>*A,?<CR>
Reply:
Interval = <Transmission interval set><CR>
Units = <Yes if Units displayed, No if units not displayed><CR>
6.6.2 N - Set Device (Transducer) Address
This command sets the address of the transducer. A valid address is a number between 0 and
32. Setting the address to zero causes the transducer to go into direct mode. All other address
settings cause the transducer to go into network mode. This command has a secondary function
and uses N,0 or *N,0 to change the length of error messages (where: 0
=
direct mode).
Command:
<SPACE>N, <New Address><CR>
Reply:
n/a
Command:
<SPACE>N,0<CR>
Reply:
n/a
Result:
Changes to short error messages.
Command:
<SPACE>*N,0<CR>
Reply:
n/a
Result:
Changes to long error messages.
In the example command below, the transducer at address 1 is changed to long error messages:
Command:
<SPACE>1:*N,1<CR>
Reply:
n/a
Query:
<SPACE>N,?<CR>
Reply:
<Device Address><CR>
Query:
<SPACE>*N,?<CR>
Reply:
Device Address = <Device Address><CR>
where: <New Address> is 0 for direct mode.
where: <New Address> is 0 for direct mode.