5.2.2 Connection of flow detector F-45-ALARM to con::cube
■
Disconnect the con::cube from power supply and
open the cover to the terminal compartment.
■
Connect the two black wires of the flow detector
as follows.
Connect the first black wire to
GND
of 12 V pow
-
er supply from con::cube (black terminals at the
bottom of the con::cube).
Connect the second black wire with the digital
input. In addition, create a cable bridge between
+ 12 VDC (red terminal at the bottom of the
con::cube) and the digital input (see figure on the
right).
■
Reconnect the con::cube to the power supply
and wait until moni::tool has started up.
Click the
Service
tab on
the moni::tool screen.
Login as
Administrator
with Password
admin1
or your individual user-
name.
Click on the icon
Digital
Inputs
.
Click on the button
Parameters
.
Click on the
+
icon
beside the input, to
which the flow detector
is connected.
1
2
3
4
5
3
2
1
4
5
micro::station / nano::station, 03-2023 Release
www.s-can.at
Copyright © s::can GmbH
19 / 36