4 Installation
4.1 Environment
!
The correct installation of measuring instruments is an important prerequisite for satisfactory operation.
Therefore the following checklist for the installation can be used to ensure that all sources for potential
operational problems can be ruled out to the greatest possible extent during the installation, allowing the
s::can monitoring system to operate properly.
Installation site:
■
Best possible weather and splash water proof protection
■
Avoid direct exposure to sunlight (see note on the panel)
■
Unadulterated, representative measuring medium
■
Measuring medium is in equilibrium state (no gas release, no precipitation,
etc.)
■
No external interferences (no electric and electro-magnetic interferences by
leakage current, earth fault of pumps, electric motors, electric power lines,
etc.)
■
Easy accessibility and availability of sufficient space (mounting, sampling,
function check, demounting)
■
Adherence to limit values (see technical specifications located at the end of
this manual)
Infrastructure:
■
Power supply for controller for operation (operational reliability, voltage, power, peak free)
■
Correct dimensioning, mounting and protection of all cables and lines (non-buckling, no risk of stumbling,
no damage etc.)
4.2
Mounting of micro::station / nano::station
For mounting and electrical installation the following tools and materials are necessary:
■
4 screws (M8) for fastening the micro / nano::station on the wall
■
Tools to open the operator terminal
■
Screw driver (3 mm) for wiring on cable terminals
■
Strap wrench to unsrew the sensor adapter of the flow cell for ISE-probe or oxi::lyser
■
Power supply cable (C-31-xx, if required)
■
Stripping tool for power supply line
■
Cable end sleeves and crimper
■
Connection fittings for inlet and outlet of micro / nano::station
!
The panels of the clean water stations are equipped with an elongated opening so that the wiring of the
sensors can also be arranged at the back of the panel. The delivery also includes 4 spacers (3 cm). From
version 2023, the back of the panel is flat.
micro::station / nano::station, 03-2023 Release
www.s-can.at
Copyright © s::can GmbH
15 / 36
Summary of Contents for scan micro station
Page 1: ...Manual micro station nano station March 2023 Release S 36 M...
Page 2: ......
Page 35: ......