PortaSonic
9000
and
components
2
/
55
UF_PS9000_BA_02_1108
2. PortaSonic 9000 and components
Illustration 1:
PortaSonic 9000 – Mounted ultrasonic sensors (below) and converter
Your PortaSonic 9000 mainly consists of the ultrasonic sensors mounted on your pipeline
system and of a converter.
2.1 Converter
The converter ensures the signal processing and makes the measurement results
available to the user.
2.2 Ultrasonic
sensors
The ultrasonic sensors are externally mounted on the pipeline and both produce and
receive the ultrasonic signals by which the flow recorded through the converter can
be determined.
Ultrasonic sensor XUC PW 20 (2 MHz), colour of housing: RED, for pipe diameters
DN10…DN100, temperatures: -40°C…150°C
Ultrasonic sensor XUC PW 10 (1 MHz), colour of housing: BLUE, for pipe
diameters: DN32…DN400, temperatures: -40°C…150°C
Ultra sensor XUC-PW 5 (0.5 MHz), color of housing: GREEN, for pipe diameters
DN200…DN6000, temperatures: -40°C…80°C (150°C optional) – depending on the
sonsor fixation (material)