Instruction 24-9428
10
Fyrite Tech
3.5.3 CO Channel
Setup
(Model 60 only)
The CO Channel Setup Screen s labeled “CO.”
1. If not already done, enter Setup per Secton 3.5.1. Repeatedly press
ENTER
untl the CO Channel Setup Screen s dsplayed.
2. Press the
or
button untl the desred parameter s dsplayed.
- no
Sensor dsabled. Use ths functon to turn off the CO chan-
nel f the sensor s mssng or known to be bad to avod
naccurate CO readngs from appearng.
- 2Ero Manual Zero. When the CO channel s set to manual zero,
the analyzer does not zero the CO sensor to ambent cond-
tons durng start up.
- A2
Auto Zero. Wth the CO channel set to auto zero, the CO
sensor s zeroed to the ambent CO level durng start up.
Important:
When using this mode the analyzer must
be turned ON in fresh air; otherwise, incorrect CO
readings will occur.
3. Press
ENTER
to move to next Setup Screen, or press
I/O
to ext setup.
3.5.4 Time Setup
(Model 60 only)
There are two Tme Setup Screens, one for hours and the other for mn-
utes. Two bars appear above the segments beng changed. The clock value
s dsplayed n a 24 hour format for setup purposes, but wll appear as
AM/PM on the prntout.
1. If not already done, enter Setup per Secton 3.5.1. Repeatedly press
ENTER
until the first Time Setup Screen is displayed—the one with
two bars over the hour dgts.