![BabyOno 613 Manual Download Page 3](http://html2.mh-extra.com/html/babyono/613/613_manual_3053021003.webp)
3
TECHNICAL SPECIFICATION
Measuring unit
°C / °F
Operating mode
Adjusted mode (body mode)
Direct mode (surface temperature) Direct mode(surface mode)
Measuring site
Forehead
Reference Body Site
Axillary
Rated output range
Body mode: 34,0 °C ~ 43,0 °C / 93,2 °F ~ 109,4 °F
Surface mode: 0 °C ~ 100 °C / 32 °F ~ 212 °F
Output Range
Body mode:34.0°C ~ 43.0 °C/ 93.2 °F - 109.4 °F
Surface mode: 0 °C ~ 100.0 °C / 32°F - 212 °F
Laboratory Accuracy
Body mode:
34.0 °C ~ 34.9 °C ± 0.3 °C/ 93.2 °F - 94.8 °F ± 0.5 °F
35.0 °C ~ 42.0 °C ± 0.2 °C / 95.0 °F - 107.6 °F ± 0.4 °F
42.1 °C ~ 43.0 °C ± 0.3 °C / 107.8 °F - 109.4 °F ± 0.5 °F
Surface
mode:
± 2 °C / ± 3.6 °F
Display resolution
0,1 °C / 0,1 °F
Three-colour backlight
(colour codes)
35.5 °C - 37.3 °C / 95.9 °F - 99,1 °F Green (normal temperature);
37.4 °C - 38.0 °C (alarm point) / 99,3 °F - 100.4 °F Yellow (slight fever)
38.1 °C - 43.0 °C / 100.6 °F - 109.4 °F Red (high fever)
Caution:
1. The colour codes only work in body temperature measuring mode.
2. The backlight is always green when measuring the temperature of surfaces.
3. In body mode 34.0-35.4°C is with green backlight.
Automatic power off time
≤ 18 s
Measuring time
≤ 2 s
Measuring distance 1 cm - 5 cm
Memory slots
50
Power supply requirements
Batteries
1.5 V (AAA) alkaline x 2 (IEC Type LR03)
Operating voltage range
2.6 V ~ 3.6 V
Ambient conditions
Operating conditions
Operating temperature: 15 °C ~ 40 °C (59 °F ~ 104 °F),
relative humidity ≤ 85%, atmospheric pressure: 70 kPa -106 kPa
Transport and storage conditions
Storage temperature: - 20 °C to + 55 °C / - 4 °F to + 131 °F,
relative humidity ≤ 93%, atmospheric pressure: 70 kPa - 106 Pa
Dimensions and weight
Weight (w/o batteries)
90 g
Dimensions
L: 138 mm X W: 95 mm X H: 40 mm
Compliance
Item
Compliance with
Device classification
Safety standard EN 60601-1: 2006+A1:2013, EN 60601-1-2: 2015
Protection type
Internally (battery) powered equipment
Type of protection Internally powerd equipment (on battery power)
Degree of protection: Non Applied part
Front panel and case labeling: EN ISO15223-1:2016
Temperature: EN ISO80601-2-56:2017
Home healthcare environment: EN 60601-1-11:2015
Safety classification of ME EQUIPMENT
Protection against electric shock Internally powered ME equipment
Applied part Non Applied part
Protection against harmful ingress of water or particulate matter IP22
Mode of operation Continuous operation
Note: Not intended to be sterillized. Not for use in an OXYGEN RICK ENVIRONMENT
Summary of Contents for 613
Page 26: ...26 1 15 30 1 2 3 4 34 0 43 0 C 93 2 F 109 4 F 1 5 1 2 3 4 3 4 1 2 AAA 4 5 30 1 2000 18 2...
Page 27: ...27 3 5 LCD 1 15 C 40 C 30 6 1 2 7 2 1 Babyono 2 Babyono 3 4 15 C 40 C 30 5...
Page 77: ...77 5 LCD 1 15 C 40 C 30 6 1 2 7 2 1 Babyono 2 Babyono 3 4 15 C 40 C 30 5...
Page 146: ...146 1 15 30 1 2 3 4 34 0 43 0 C 93 2 F 109 4 F 1 5 1 2 3 4 3 4 1 2 AAA 4 5 30 1 2000 18 2 3 5...
Page 147: ...147 1 15 C 40 C 30 6 1 2 7 2 1 Babyono 2 Babyono 3 4 15 C 40 C 30 5...
Page 176: ...176 1 15 30 1 2 3 4 34 0 43 0 C 93 2 F 109 4 F 1 5 1 2 3 4 1 2 AAA 4 5 30 1 2000 18 2 3 5...
Page 177: ...177 1 15 C 40 C 30 6 1 2 7 2 1 Babyono 2 Babyono 3 4 15 C 40 C 30 5...
Page 182: ......
Page 183: ......
Page 184: ...Cat No 613...