Mod.BM43
COM
Vers.1.0
16/11/2005
Page 9 of
95
B.M. Tecnologie Industriali
Via Praimbole 13
35010 LIMENA (PD)
VAT No. IT02459940280
Tel. +39 (0) 49 884.16.51
Fax +39 (0) 49 884.16.54
E-Mail [email protected]
Web www.bmtecnologie.it
1.4 INTEGRATED REAL TIME CLOCK (RTC)
The user’s settings are saved into the device’s memory for at least 2100 years in case of
power failures. The device could be protected by a PASSWORD.
The RTC remains active until the battery discharge tension is 1.5 V.
Ref. to menu M60 for resetting current date and time in the format: yy, mm, dd.
1.5 PRODUCT ID
Each device has a serial number, generated during its production. It could not be changed,
it should be used in case of repair and it could be displayed by menu M61.
1.6 TECHNICAL FEATURES
FEATURE
SPECIFICATION
Linearity
0.5%
Repeatability
0.2%
Accuracy
+/- 1% of the reading value > = 0.2 m/s
Response time
From 0 to 999 seconds, set by the user.
Speed
+/- 32 m/s
Pipes diameter
From DN20 to DN 6000
Eng. Units
Meters, Feet, Cubic meters, Cubic feet, USA
Gallons, Imperial Gallons, USA Million
Gallons, set by the user.
Totalizers
7 digit for positive, negative and net
flowrate.
Measurable liquids
All the liquids (virtually)
Safety
Possible to set a password for blocking the
device.
Display
Graphic dislya4 lines, 16 characters.
Interface
RS232-C from 75 to 57600 BPS.
Transducers
TS-2, TM-1, TL-1, or High temperature
sensors, at customers request.
Cable lengths
From 2 x 5 m up to 2 x 500 m
Supply
3 x AAA Ni-MH batteries (included) for up
to 10 hours of continuous operation
Data logger
Internal data logger to save up to 2000
lines of data.
Manual totalizer
7 digit totalizaer for manual acquisitions
and calibrations.
Housing material
ABS
Housing dimensions
460 (L) x 400 (W) x 110 (H) mm
Total weight
4.5 kg batteries included