Hot surface warning

 

      

 Under normal conditions, the temperature of glass surface of UV transilluminator is below 

50 °C and safe to touch. However, if the system malfunctions, it is possible that the 

glass surface temperature exceeds 80°C. Please exercise caution when touch the glass 

surface with hand when this occurs.

Laser Safety

General Information
Azure Biosystems imaging system models c600 and c500 both include a laser illumination system that 

provides 660nm and 785nm Narrow band excitation for NIR applications. The imaging systems are 

certified to comply with CE, UL and CSA safety standards and consists of two fully enclosed electronic 

modules located on the right and left interior sides of the imaging chamber. This laser system is by its 

appropriate classification and definition a non-removable laser system as it is not operable when the 

laser modules themselves are removed from within the system.
Safety Features
Azure Biosystems imaging systems are designed to prevent direct and collateral human exposure to 

laser radiation by means of a safety interlock switch located on the right front side of the imaging 

chamber. The safety interlock reacts to “door open” and “door closed” states and defeats all power 

to  internal  light  sources  when  the  door  is  in  the  “open  position”.  Lasers  and  other  internal  system 

light sources will not power on unless the door is fully closed. If the access door is opened during 

imaging, all light sources will immediately power off to prevent human exposure to internal illumination 

sources. Right and left laser modules located in the c500 and c600 imaging chamber are installed in a 

highly diffused protective metal enclosure preventing exposure to full laser radiation power. In addition, 

the entire laser system is fully enclosed within the system enclosure and there are no viewing ports, 

windows, or openings to facilitate viewing of, or exposure to laser radiation fields from direct impact, 

reflection, or leakage. 
Maintenance
Azure Biosystems c500 and c600 laser systems do not require regular, periodic, or preventive maintenance 

in the form of adjustments, calibrations, cleaning, or other standard maintenance procedures to maintain 

optimal performance, thereby removing the need for users or their service technicians to initiate any 

actions where exposure to laser radiation would occur. 
Serviceability
Replacement of faulty laser modules is a manufacturer-only repair action and not a customer-service 

action.  Laser  repair  or  replacement  may  be  performed  in  the  field  by  Azure  Biosystems  authorized 

service technicians, or by return of the entire system to Azure Biosystems, or its authorized service 

location(s) for laser repair or replacement. Lasers are deemed to be faulty or defective if users discover 

images that show evidence of output signal level loss in either the left or right laser module, a significant 

difference between the output signal levels between the left and right lasers modules, or complete loss 

of output signal level in either or both laser modules. Users or their service technicians should make no 

attempt to determine the cause of faulty laser operation, and should promptly contact Azure Biosystems 

at 

[email protected]

 or their nearest Azure Biosystems authorized service location.

Summary of Contents for c150, c200, c300, c400, c500, c600

Page 1: ...Azure cSeries User Manual Page 1 Azure cSeries User Manual for the c150 c200 c300 c400 c500 and c600 NIR RGB CHEMI WHITE LIGHT BLUE LIGHT UV Rev 20150428...

Page 2: ...coming in through the sides or around the lenses Always cover all skin that may be exposed to UV light especially the face neck hands and arms Always make sure that any UV protection devices such as t...

Page 3: ...al illumination sources Right and left laser modules located in the c500 and c600 imaging chamber are installed in a highly diffused protective metal enclosure preventing exposure to full laser radiat...

Page 4: ...however are specifically excluded Defects caused by improper operation Repair or modification done by anyone other than Azure Biosystems or the company s authorized agent Use of spare parts supplied b...

Page 5: ...3 2A Imaging Workflow 9 3 2B Band Excision and Light Safety Switch 10 3 3 Chemiluminescent Applications c300 systems and above 10 3 3A Imaging Workflow for Chemiluminescent Applications 11 3 4 RGB App...

Page 6: ...5nm trans illumination It is mounted on a pull out tray EPI Blue and White LEDs The cabinet also contains EPI white and blue light panels on each side of the cabinet interior walls EPI RGB LEDs High i...

Page 7: ...ter solvents or corrosive materials on a flat and stable surface with adequate clearance on all sides The system must remain stationary during operation Further the system should be placed away from i...

Page 8: ...ng screen during initialization Once the system is done initializing the software displays the gallery tab and is ready to use 2 5 Attach Other USB Input Devices to the System You may attach regulator...

Page 9: ...ll display your image with dark and light pixels reversed SATURATION Selecting Saturation will show you images with intensities beyond the dynamic range of the camera SINGLE MULTICHANNELVIEWING When a...

Page 10: ...ht source for gels that are stained with dyes that excite in the blue range Visible For dyes that are visible without additional excitation use the Visible light source When selected the software will...

Page 11: ...he resolution of the camera and image At the Lowest sensitivity setting you will be taking a full resolution image with the camera And the Highest sensitivity setting the resulting image will be 0 3 M...

Page 12: ...amera Recommendation Set the exposure time to 30 Seconds by hitting 3 then 0 and then S Keep the Sensitivity setting on the default setting NORMAL Select CUMULATIVE After hitting CAPTURE a series of i...

Page 13: ...e time 5 Select CAPTURE The channels will be captured sequentially 6 The resulting image will be displayed in the gallery tab 3 5 NIR Applications c500 and c600 For all NIR blots use the Blot Tray on...

Page 14: ...d like them assigned to Set the exposure time for each channel 3 Select the aperture setting the smaller the number the wider the aperture For chemiluminescence we recommend 0 95 for fastest light gat...

Page 15: ...CUS CALIBRATION Allows the technician to reset the focus setting in case of a shift in focus CREATE FLATS Allows the technician to recreate flat calibration files 5 Help Selecting the help tab will al...

Page 16: ...ower Off Restart the Tablet To turn off the system shut down the tablet by a Swiping from the right to left on the right side of the screen to bring up the right side toolbar b Select SETTINGS c Selec...

Page 17: ...st of programs to uninstall select CSERIES CAPTURE SOFTWARE 4 Once it is uninstalled you can begin installing the new software 5 Select the SETUP file in the new software folder 6 Select RUN from the...

Page 18: ...the default 12 If it asks you to replace the existing cSeries folder select YES then INSTALL 13 On the CP210x USB to UART Bridge Driver Installer screen select NEXT 14 On the next screen accept the ag...

Page 19: ...the software and make sure the hardware is turned on and re launch the software Copyright 2013 Azure Biosystems Inc All rights reserved The Azure Biosystems logo and Azure are trademarks of the Compan...

Reviews: