3.3

A

 Imaging Workflow for Chemiluminescent Applications 

 

1.  Remove the Blot Tray from the cSeries door. 

 

2.  After you have applied the substrate to your blot, place your blot in the very center of the tray. 

 

3.   Place the blot tray in one of two positions: on the UV transilluminator, or on the chemi shelf, in 

the middle of the imaging system. 

 

  •  Use the position on the UV transilluminator for larger blots. 

 

  •   Use the position on the chemi shelf for small blots. Using this top shelf increases the 

resolution of the blot. 

 

4.  Close the door to the system. 

 

5.   Select an exposure time. Choose the exposure time manually or use autoexposure which will 

allow the software to capture image for you using the full dynamic range of the camera.

 

  •   Recommendation: Set the exposure time to 30 Seconds by hitting 

3

 then 

0

 and then 

S

Keep the Sensitivity setting on the default setting, 

NORMAL

. Select 

CUMULATIVE

. After hitting 

CAPTURE

, a series of images will be displayed, and you can select the best one from the 

gallery tab. 

 

  •  Maximum exposure time is 60 minutes. 

  6. Select 

CAPTURE

 

7.  The images will be displayed in the gallery tab.

3.4 RGB Applications (c400 and c600) 

Selecting 

RBG

 will give you the follow screen:

The RGB setting is for fluorescent Western blots labeled with 

Cy5/Cy3/Cy2 or dyes with similar excitation and emission 

spectrums. 
For all RBG blots, use the Blot Tray on the UV transilluminator 

position only. The blot tray provides a dark background to 

improve the contrast on the blot image. Placing this Tray on 

the UV transilluminator position provides even illumination 

for your samples.

Positions for blot tray:

Chemi Shelf (for chemi blots only)

UV Transilluminator (for all blots)

Summary of Contents for c150, c200, c300, c400, c500, c600

Page 1: ...Azure cSeries User Manual Page 1 Azure cSeries User Manual for the c150 c200 c300 c400 c500 and c600 NIR RGB CHEMI WHITE LIGHT BLUE LIGHT UV Rev 20150428...

Page 2: ...coming in through the sides or around the lenses Always cover all skin that may be exposed to UV light especially the face neck hands and arms Always make sure that any UV protection devices such as t...

Page 3: ...al illumination sources Right and left laser modules located in the c500 and c600 imaging chamber are installed in a highly diffused protective metal enclosure preventing exposure to full laser radiat...

Page 4: ...however are specifically excluded Defects caused by improper operation Repair or modification done by anyone other than Azure Biosystems or the company s authorized agent Use of spare parts supplied b...

Page 5: ...3 2A Imaging Workflow 9 3 2B Band Excision and Light Safety Switch 10 3 3 Chemiluminescent Applications c300 systems and above 10 3 3A Imaging Workflow for Chemiluminescent Applications 11 3 4 RGB App...

Page 6: ...5nm trans illumination It is mounted on a pull out tray EPI Blue and White LEDs The cabinet also contains EPI white and blue light panels on each side of the cabinet interior walls EPI RGB LEDs High i...

Page 7: ...ter solvents or corrosive materials on a flat and stable surface with adequate clearance on all sides The system must remain stationary during operation Further the system should be placed away from i...

Page 8: ...ng screen during initialization Once the system is done initializing the software displays the gallery tab and is ready to use 2 5 Attach Other USB Input Devices to the System You may attach regulator...

Page 9: ...ll display your image with dark and light pixels reversed SATURATION Selecting Saturation will show you images with intensities beyond the dynamic range of the camera SINGLE MULTICHANNELVIEWING When a...

Page 10: ...ht source for gels that are stained with dyes that excite in the blue range Visible For dyes that are visible without additional excitation use the Visible light source When selected the software will...

Page 11: ...he resolution of the camera and image At the Lowest sensitivity setting you will be taking a full resolution image with the camera And the Highest sensitivity setting the resulting image will be 0 3 M...

Page 12: ...amera Recommendation Set the exposure time to 30 Seconds by hitting 3 then 0 and then S Keep the Sensitivity setting on the default setting NORMAL Select CUMULATIVE After hitting CAPTURE a series of i...

Page 13: ...e time 5 Select CAPTURE The channels will be captured sequentially 6 The resulting image will be displayed in the gallery tab 3 5 NIR Applications c500 and c600 For all NIR blots use the Blot Tray on...

Page 14: ...d like them assigned to Set the exposure time for each channel 3 Select the aperture setting the smaller the number the wider the aperture For chemiluminescence we recommend 0 95 for fastest light gat...

Page 15: ...CUS CALIBRATION Allows the technician to reset the focus setting in case of a shift in focus CREATE FLATS Allows the technician to recreate flat calibration files 5 Help Selecting the help tab will al...

Page 16: ...ower Off Restart the Tablet To turn off the system shut down the tablet by a Swiping from the right to left on the right side of the screen to bring up the right side toolbar b Select SETTINGS c Selec...

Page 17: ...st of programs to uninstall select CSERIES CAPTURE SOFTWARE 4 Once it is uninstalled you can begin installing the new software 5 Select the SETUP file in the new software folder 6 Select RUN from the...

Page 18: ...the default 12 If it asks you to replace the existing cSeries folder select YES then INSTALL 13 On the CP210x USB to UART Bridge Driver Installer screen select NEXT 14 On the next screen accept the ag...

Page 19: ...the software and make sure the hardware is turned on and re launch the software Copyright 2013 Azure Biosystems Inc All rights reserved The Azure Biosystems logo and Azure are trademarks of the Compan...

Reviews: