54
•
Specifications
Manual External Command Mode
When active the following readings are disabled:
Micropipette tip resistance, current and power.
Rear panel
All connectors grounding wired to internal ground, and to DC power supply
negative terminal. DC power supply output isolated from mains ground.
Factory Setup (DB-9F connector)
Serial connection to PC for firmware upgrades and factory testing only. Must
be left disconnected during normal operation.
AP-1A Headstage (DB-15F connector)
Used only for AP-1A-1MU and AP-0.1A-0.1MU headstages.
(
Connect to or disconnect from the main unit only when the power is off.
)
Outputs
I
output
(micropipette current)
Sensitivity = 1 V/
µ
A for AP-1A-1MU headstage
Sensitivity = 0.1 V/
µ
A for AP-1A-0.1MU headstage
Output impedance = 500
Ω
V
output
(voltage applied to the micropipette)
Sensitivity = 0.1 V/V
Output impedance = 475
Ω
Sync Output
Zero to 5 V signal, with the same width as the generated pulse (including
the approach waveform), output impedance 10 k
Ω
. During bi-level pulse
Axoporator 800A Theory and Operation, Copyright 2005 Axon Instruments / Molecular Devices, Corp.