15.7 Function for setting serial ports mode and parameters (Port)
After choosing function from menu
we have two options:
-
Port 1
-
Port 2
By choosing equivalent port we can
change parameters of this port. If the
scale is equipped with only one serial
port, choose
Port 1
option.
Turning on the function enables to
set
the
following
transmission
parameters
(standard
parameters
underlined:
-
transmission speed (
bAud: 1200,
2400, 4800, 9600, 19200, 38400,
57600, 115200
),
-
the number of bits in a byte (
bit
:
7,
8
),
-
parity control (
PArItY
:
0, 1; Odd: 0,
1
),
-
data transmission protocol (
Prot:
Long, Short, Farb, Fis_E, Fis_A,
Pr_CC, Eltron, out),
-
ustawienie
trybu
pracy
portu
szeregowego
SendInG.
After choosing
SendInG
-
butt. P
– transmission after using
key without stabilization,
-
b. P Stb
– transmission after using
key and indication stabilization
-
Auto
– after taking on and off the
load without using
key,
-
Cont.
- about 10 results per second,
-
Out.
Auto
is suggested when working with
printer.
To set desired transmission parameters activate
Port
function, choose proper port 1
or 2 and then choose appropriate parameter and press
T
key to accept needed
parameter value. The example at the left presents how to set transmission speed
value to 9600bps.
0.00g
Port
baud
-9600
-4800
bIt
out
Odd
PArItY
MENU
T
T
T
T
T
T
T
Port-1
T