Steps during calibration with external weight:
Press
MENU
key to display user
functions, shown one by one in loop.
Choose calibration function pressing
T
key when
CALIbr
function
appears.
The
following
options
will
be
displayed:
-
CAL oFF
– switch currently carried
out internal calibration off
-
CAL on
– perform calibration with
external weight
-
CAL Prn
– printout of calibration
report
-
CAL tM
– set time interval for internal
calibration
-
CAL
C
- set temperature difference
for internal calibration
When
CAL on
option appears press
T
key to select function for
calibration with external weight.
Press
key several times to select
calibration weight value, which will be
used for calibration.
Confirm pressing
T
key.
When
LOAD
message appears put
calibration weight on the pan and press
T
key.
Wait until calibration process is
finished.
When
unLOAD
message appears take
calibration weight off.
Wait for end of balance zeroing.
Wait until internal calibration is
finished.
Balance is ready to work.