![Axis AGN/C Series User Manual Download Page 22](http://html.mh-extra.com/html/axis/agn-c-series/agn-c-series_user-manual_3038948022.webp)
22
USER MANUAL
__________________________________________________________________________________________________________________
15.5 Percent function (PErCEnt)
The
function
allows
displaying
weighing result in percents.
Measurement is performed in two
phases:
-first phase – weighing reference
mass (mass referenced to 100% )
-second phase – weighing sample
mass as a percent of reference mass
Weighing result is displayed in
various
formats,
depending
on
reference mass value. For reference
mass values between 0÷3,5% of
balance capacity, format of weighing
result is
100.0
, for values between
3,5÷35% it is
100.00
and above 35%
-
100.000
.
The function has the following
options:
-
PEr oFF
– deactivate the function,
-
PEr on
– save current indication as
100%, start weighing in percents,
-
out –
exit without changes.
In order to go back to indication in
mass units use
key. Using this
key once more will cause return to
displaying in %.
Caution:
1.
PEr Err
message informs that
reference mass is less than
0,5
Min or was not defined.
2.
When the balance shows weighing
result in percents,
T
key
works as usual.
0.0000g
19.07 g
PErCEnt
PEr oFF
PEr on
100.00%
50.00%
100%
50%
MENU
T
T