Operator Manual, Supercharger 60
R3.0
Page 16 of 62
3.3.2.4.
NOTE: The colon flashes at the rate of one cycle per second regardless
of the selected mode.
3.3.3.
ACCURACY AND STABILITY:
3.3.3.1.
±0.01%
3.3.4.
POWER FAILURE PROTECTION:
3.3.4.1.
An internal rechargeable battery maintains (for several hours) the clock
and other vital circuits for a dependable resumption of operation after a
power failure.
3.3.4.2.
The colon will be lit in case of a power failure.
3.3.4.3.
NOTE: Do not operate the Charger-Analyzer without this battery or
with a battery in bad condition. Replace only with the same type, 7 or 8
cells (8.4V or 9.6V), rechargeable).
3.3.5.
TIME LIMITS:
3.3.5.1.
0 to 9 in Main
3.3.5.2.
0 to 79 in Total
3.3.5.3.
NOTE: The time limit in Total is dependent on the type of Timer.
Newer instruments, characterized by a Timer Speed Switch mounted
on the Front Panel have a limit of 79 (hours or minutes) while older
units have a limit of 60.
3.4
LINE INPUT VOLTAGE:
3.4.1.
115V(*), 208V, 230V, 245V
Note: For proper operation of this instrument, power line voltages must be
within ±10% of the nominal.
3.4.2.
The instrument is normally wired for 230V operation.
3.4.3.
A terminal block at the transformer inputs allows field re-wiring for 245V, 208V
or 115V operation.
See figs 3, 4 and 5
.
3.4.4.
For operation at line voltages in excess of 220V at 50Hz, it is required that the
245V tap be used.
3.4.5.
NOTE (*): Operation at 115V is not recommended at battery charging currents
above 15A due to the increased line current.
Summary of Contents for Supercharger 60
Page 7: ...Operator Manual Supercharger 60 R3 0 Page 7 of 62 Figure 1 Block Diagram ...
Page 20: ...Operator Manual Supercharger 60 R3 0 Page 20 of 62 Figure 2 Front Panel ...
Page 44: ...Operator Manual Supercharger 60 R3 0 Page 44 of 62 Figure 6 Meters Board Adjustments ...
Page 52: ...Operator Manual Supercharger 60 R3 0 Page 52 of 62 Figure 11 Circuit Board Sequence ...
Page 57: ...Operator Manual Supercharger 60 R3 0 Page 57 of 62 Figure 13 Location of Current Limiters ...
Page 58: ...Operator Manual Supercharger 60 R3 0 Page 58 of 62 Figure 14 Power Block Diagram ...