background image

3.5 Exiting calibration

To exit calibration mode and return to normal operation use the up arrow key as follows:
• If the positioner is at menu level in the calibration, as determined by LCD displaying a menu name

only (MCAL, etc.), press the up arrow key once to exit CAL mode.

• If the positioner is at function level in the calibration, as determined by LCD displaying a function

and menu name only (MCAL Lo, etc.), press the up arrow key once to enter the menu level and
once more to exit CAL mode.

• When the calibration mode is exited the menu and function names will no longer be displayed by

the LCD. The LCD will be displayed ‘OK’.

Exiting can not be done during a calibration procedure. When a calibration function is initiated, the 
user must wait until the function’s calibration is complete before being able to exit calibration.
The up arrow key can be used, as described above, to move to the Menu level and then to exit 
CAL mode.

3.6 Manual override of input signal (via on-board keypad)

The positioner has a feature which allows the operator to override the analog signal and change 
valve position of the SmartCal. This is done from the Stro (manual override stroke menu). Enter 
calibration as described in section 3.1 and use the down arrow button to cycle to the Stro menu. 
Enter this menu and control the position of the valve as shown below.

Up arrow

(exit calibration)

Down arrow

(from previous menu)

Down arrow

(to next menu)

Down 

arrow

Down 

arrow

Down 

arrow

UP

UP

UP

Slow 

move

Press & 

hold up 

or down 

arrow 

valve will 

move 

0.1%

Valve 

full open

Use up 

& down 

can 

move 

valve 

slowly

Fast 

move

Press up 

or down 

arrow 

one time 

valve 

moves 

5%

Valve 

closed

Use up 

& down 

can 

move 

valve 

slowly

AVID SmartCal valve positioner

Installation & operating instructions

page 14

Summary of Contents for SmartCal Series

Page 1: ...ctions 22 4 5 Print 23 5 Trouble shooting 24 5 1 Preliminary checks 24 5 2 FAQ 25 6 Specifications 26 7 Error codes 26 8 Exploded parts list 27 Appendices A Procedure to adjust Err 3 setting 28 B Proc...

Page 2: ...ilure condition occurs an error message is displayed on the local liquid crystal display 1 2 Principle of operation Unlike conventional positioners the SmartCal positioner feeds back valve position wi...

Page 3: ...d sensors The SmartCal positioner has the capability to monitor its operation If an error or failure condition occurs it will be displayed on the local LCD through the HART interface and displayed on...

Page 4: ...of the positioner Actuator in fail position ModMount Supply Port 1 Port 2 TURNS COUNTER CLOCKWISE from fail position TURNS CLOCKWISE from fail position 2 1 Mounting positioner on a rotary actuator Co...

Page 5: ...oner sensor TURNS COUNTER CLOCKWISE From fail position TURNS CLOCKWISE From Fail Position 2 2 Mounting remote positioner on a rotary actuator Condition 1 Actuator fails in a clockwise direction Turns...

Page 6: ...y unscrewing 2 mounting screws Wire the positioner sensors back to the positioner using the cable provided see Figure 2 3 4 Conductor cable cut to required length and run through conduit M20x1 5 1 2 N...

Page 7: ...embly to the stem of the actuator A coupler block normally is needed to extend the magnet assembly outside the yoke area and into the sensing range of the magnetic pick up unit Step 2 Fasten the mount...

Page 8: ...n the actuator s yoke that lines up with the red arrow on the magnet assembly 3 Place a third mark on the yoke centered between the upper and lower limit marks 4 Lastly mount the positioner to the bra...

Page 9: ...ribed in Section 3 3 Step 1 Step 2 Mount the position sensor housing so that the conduit entry faces away from the diaphragm or cylinder See Figure 2 5 A Step 3 Mount positioner at a remote location N...

Page 10: ...to decrease in outlet port 2 of the positioner Note Air supply to the positioner must be clean dry oil free instrument air 5 microns per ISA S7 3 and ISO 8573 1 Maximum supply pressure is 8 3 bar All...

Page 11: ...ally connected suitable to the rated data via a certified isolating interface zener barrier placed outside the hazardous areas 4 For ambient temperatures below 10 C and above 60 C use field wiring sui...

Page 12: ...menu Down arrow to next menu Up arrow Up arrow Down Down Down Down Down Down Down Down Down Quick opening flow Equal percentage flow Rotary valve Fail open Use up down arrows to select open speed 01...

Page 13: ...row Down arrow Press CAL key Down arrow Down arrow Down arrow Down arrow Down arrow Down arrow Down arrow Down arrow Down arrow Set mA for zero position Set mA for span position Sensor calibration Tra...

Page 14: ...s complete before being able to exit calibration The up arrow key can be used as described above to move to the Menu level and then to exit CAL mode 3 6 Manual override of input signal via on board ke...

Page 15: ...Sensor calibration 5 Trnd Transducer calibration 6 mA Milliampere calibration 7 Xmr Transmitter calibration Menu 3 Cofg configuration Entering this menu allows you access to the following five configu...

Page 16: ...libration is a self adjustment that sets the positioners Hall effect circuitry This is automatically done during the ACAL automatic calibration routine The sensor calibration also shows up under the M...

Page 17: ...1 which is equivalent to a deadband of 0 5 The value 20 highest deadband value has a range of 20 which is equivalent to a deadband of 10 LCD This feature configures LCD timeout The range is 1 to 60 m...

Page 18: ...or the connection to the 4 20 mA controls see the schematic in appendix D The SmartCal DTM can be used with several FDT frame applications If you don t have any you can use the PACTware configuration...

Page 19: ...D values or tune them after auto calibration The positioner is configured for 0 100 control By changing the parameters in the Travel Limit Setting menu you can alter the lower and upper limits Only a...

Page 20: ...alve movements The end result is the number of full open close cycles The value Accum DB is the minimum stroke that can be counted If the counted value exceeds this limit the alarm will be activated T...

Page 21: ...ware optional This facilitates the comparison of curves and the preparation of reports 4 3 Diagnosis Selection of Diagnosis produces an overview of all SmartCal alarms and statuses It shows proper ope...

Page 22: ...on Capture Analog Digital and Digital tuning Each selection has its own menu Please follow these instructions in order to carry out the appropriate calibration The Analog selection uses the mA signal...

Page 23: ...e test procedure will be cancelled and a full auto or manual calibration has to be carried out Note while executing certain functions e g calibration communication between pc and SmartCal may stop and...

Page 24: ...ic connection Single acting Output port 1 should be piped to drive the actuator away from the valves fail position Output port 2 should be plugged See Section 2 6 Double acting Output port 1 should be...

Page 25: ...linear is desired enter calibration and make this change See Calibration instructions Section 3 5 After removing power to the positioner there is full pressure to output port 1 and zero pressure to o...

Page 26: ...gnostics HART protocol software utilizing HART protocol AMS or FDT DTM Enclosure Material Engineered resin Class of equipment NEMA type 4 4x or IP66 Weight 3 3 Kg Air connections 1 4 NPT or BSP Std fl...

Page 27: ...assembly 3 1 Electronics module assembly 4 1 Transducer assembly 5 1 Housing assembly 6 1 Manifold assembly 7 1 Direct mount assembly 8 Exploded parts list Dimensions mm Front view ModMount Namur pat...

Page 28: ...e Error 3 for an explicit pressure value loosen the lock nut on the adjustment screw and gently turn the screw clockwise as far as it will go Do not force the screw past its limit or the Error 3 diaph...

Page 29: ...ing and ground cable Remove the screw that secures ground cable See Figures below 3 Remove the canister with electronics from SmartCal enclosure Canister cover Indicated screw secures cover and housin...

Page 30: ...LCD will display a failure message and the transmitter output if so equipped will go to pre determined current to indicate the failed state The transmitter output can be configured to go to 3 4 mA or...

Page 31: ...he 4 20 mA signal generator Note The transducer operates between 0 and 3 3 mA Therefore make sure when turning on the current supply s power the current is turned down within this range Applying a cur...

Page 32: ...ntain optimum positioner performance To clean the spool valve the spool piece needs to be removed see diagram below Prior to removing the spool make sure the positioner is out of service and all air p...

Page 33: ...d twisted pair Belden 8762 or equivalent Maximum distance from HART Multiplexer to positioner is 1800 m 6000 feet Shield shall be connected to ground at one point only in order to avoid ground loops a...

Page 34: ...ram Supply air Pilot air assembly Outlet port 2 Outlet port 1 20 micron air filter Air to actuator air out of transducer Air to transducer Pilot air transducer page 34 AVID SmartCal valve positioner I...

Page 35: ...ed apparatus used in an approved configuration such that A SmartCal Vmax Voc and Vt of associated apparatus B SmartCal Imax Isc and It of associated apparatus C Ci of SmartCal cable capacitance Ca of...

Page 36: ...riers may be used where both channels have been certified for use together with combined entity parameters The following conditions must be satisfied Voc or Vo Vmax or Ui Ca Ci C Cable Isc or Io Imax...

Page 37: ...sociated apparatus E In cases where the Rosemount 275 HART Communicator is connected between the associated apparatus and the SmartCal cable inductance should be determined in accordance with Rosemoun...

Page 38: ...intrinsically safe associated apparatus ground 4 Rosemount 275 HART Communicator is NOT FMRC approved for use in Class II and III Hazardous Locations 5 For Class II and III locations where rigid metal...

Page 39: ...pparatus ground 4 Rosemount 275 HART Communicator is NOT FMRC approved for use in Class II and III Hazardous Locations 5 For Class II and III locations where rigid metal conduit is not used seal Smart...

Page 40: ...r Vo Vmax or Ui Ca Ci C Cable Isc or Io Imax or Ii La Li L Cable 2 Associated apparatus manufacturer s installation drawing must be followed when installing this equipment 3 Control equipment connecte...

Page 41: ...Use the down arrow to change No to Yes To begin Factory reset procedure press the Cal button 4 When mA METER CALIBRATION appears release the Up arrow button 5 After releasing the Up arrow button you w...

Page 42: ...witch Alert 1 OpnLmtSw Alert 2 ClsLmtSw Alert Basic Config 1 Valve Type 2 Flow Type 3 Display Mode 4 Control Mode Cal Config 1 PID Parameters 2 Slew Speed 3 Setpoint Capture 4 Limit settings 5 Digital...

Reviews: