4
Pentamidine is not an approved medication for use with this device
.
Always dispose of any remaining medication in the medication cup after
each use. Use fresh medication each time you use the device.
Do not leave the device or its parts where it will be exposed to extreme
temperatures or changes in humidity, such as leaving the device in a
vehicle during warm or hot months, or where it will be exposed to direct
sunlight.
Do not use or store the device where it may be exposed to noxious fumes
or volatile substances.
Make sure that the nebulizer kit is clean before use.
Do not use in anaesthetic or ventilator breathing circuits.
Not suitable for use in the presence of flammable anesthetic mixture with
air or oxygen or nitrous oxide.
Do not operate where oxygen is being administered in a closed
environment (such as an oxygen tent).
Provide close supervision when this device is used by, on, or near infants,
children or compromised individuals.
If the device is used continuously, the service life of the device may be
shortened.
Limit use to 30 minutes at a time, and allow a 30 minutes interval before
using the device again.
When using the device, the main unit may become hot.
Do not touch the main unit for other than necessary operation such as
turning off the power while nebulizing.
Do not insert any object into the compressor.