29
The AH-C10 is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customers or the
users of the AH-C10 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the AH-C10 as recom-
mended below, according to the maximum output power of the communi-
cations equipment.
Output Power
of the Transmitter
in Watt
Separation distance according to frequency
of transmitter in meter
150kHz to 80MHz
d =1.2 P
80MHz to 800MHz
d =1.2 P
800MHz to 2.5GHz
d =2.3 P
0.12
0.12
0.23
0.37
0.37
0.74
1.17
1.17
2.33
3.69
3.69
7.38
0.01
0.1
1
10
100
11.67
11.67
23.33
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P
is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note: At 80MHz and 800MHz, the separation distance for the higher
frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects, and people.
√
√
√
Recommended separation distance between portable and mobile RF
communications equipment and the AH-C10