20
·
Do not carry or leave the nebulizer kit with medication in the
medication cup.
·
Do not wrap the power cord around the compressor (main unit).
CAUTION
PROBLEM
CAUSE
SOLUTION
No power to unit
when the power
switch is on.
No nebulization
or low nebuliza-
tion rate when
the power is on.
The power cord is not plugged into
an electrical outlet.
Turn the power switch off. Plug the power cord
into an electrical outlet. Turn the device on.
The device is
abnormally loud.
The device is
very hot.
The air filter cover is incorrectly
attached.
Attach the air filter cover correctly. Make sure
the air filter cover is not blocked.
The compressor is covered.
Do not cover the compressor with any type of
cover during use.
Operating continuously over 30
minutes.
Limit use to 30 minutes at a time and allow a
30 minutes interval before using the device
again.
The nozzle is blocked.
Clean and disinfect the nebulizer kit to remove
the blockage.
The nebulizer kit is tilted at an
incorrect angle.
Hold the nebulizer kit correctly. Do not tilt the
nebulizer kit so the angle of the kit is greater
than 45 degrees.
No medication in the medication
cup. Too much or too little
medication in the medication cup.
Add the correct amount of prescribed
medication to the medication cup.
The nebulizer kit is not correctly
assembled.
Make sure the nebulizer kit is correctly
assembled and the inhalation accessory is
correctly attached.
The air tube is incorrectly attached.
Make sure the air tube is correctly
attached to the compressor and the
nebulizer kit.
The air tube is folded or damaged.
The air tube is blocked.
Make sure the air tube is not folded, kinked or
bent. Inspect the air tube for any damage.
Replace the air tube if damaged.
The air filter is dirty.
Replace the air filter with a new clean air
filter.
Troubleshooting Guide
Troubleshooting