background image

NE-C28P-E_C_M05_151217.pdf

16E1107

EN

Intended use

Medical Purpose

The intended purpose of this product is:

- To allow for nebulization treatment of respiratory tract disorders.

- To remove mucus form the nasal cavities to ease breathing.

Intended User

• Patient and/or caregiver under the direction of legally certified medical 

experts, such as doctor, nurse and therapist or qualified medical experts 

for home treatment.

• The user should also be capable of understanding general operation of 

DuoBaby and the content of this instruction manual.

Intended Patients

Persons from the age of 1 month who suffer from upper airway conditions 

(such as blocked nose, cold, allergies) and/or lower respiratory conditions 

(such as asthma, bronchitis, bronchiolitis).

Environment

This product is intended for use in a general household.

Durable period

Durable periods are as follows, provided the product is used to nebulize 

3 times a day for 10 minutes each time at room temperature (23°C). 

Durable period may vary depending on usage and environment.
 

Compressor (Main unit) 

5 years

 

Nebulizer Kit 

1 year

 

Mouthpiece 

1 year

 

Air Tube 

1 year

 

Air Filter 

70 applications

 

Baby Mask 

1 year

 

Nasal Aspirator 

1 year

Precautions for use

Warnings and cautions described in the instruction manual should be 

observed.

Know your unit

Important safety instructions

 

Warning:

Indicates a potentially hazardous situation which, if not avoided, 

could result in serious injury.

• A single-patient use is recommended.

• Do not use the device for other use than nebulization for human aerosol 

therapy and for removing the mucus.

• Manufacturer will not be held responsible for inappropriate use of the 

device.

• Use only original OMRON accessories as indicated in this instruction 

manual.

• The correct functioning of the device may be affected by electromagnetic 

interference caused by malfunctioning television sets, radios, etc. If this 

occurs, move the source of interference away from the device or use a 

different plug.

• For type, dose, and regime of medication follow the instructions of your 

doctor or respiratory therapist. Before use check that you are using the 

correct Inhalation top.

• If you feel or see anything unusual during use, stop using the device 

immediately and consult your doctor.

• Do not use only water in the nebulizer for inhaling purposes.

• Keep the device out of the reach of unsupervised infants and children. 

The device may contain small pieces that can be swallowed.

• In view of their length, the power cord and air tube could constitute a 

strangulation hazard.

• Do not store the air tube while there is moisture remaining inside.

• Always dispose of any remaining medication after use, and use fresh 

medication each time.

• Do not use in anaesthetic or ventilator breathing circuits.

• Never block the ventilation slots. Never place the device where the 

ventilation slots may be obstructed during operation.

• Never leave the nebulizer kit with medication near the ventilation slots, 

this is to avoid medication entering the device.

• Store the parts in a clean location for avoiding infection.

(Risk of electrical shock)

• Never plug or unplug the device with wet hands.

• The compressor and power cord are not waterproof. Do not spill water or 

other liquids on these parts. If liquid does spill on these parts,immediately 

unplug the power cord and wipe off the liquid.

• Do not immerse the compressor in water or other liquid.

• Do not use or store the device in humid locations.

• Do not operate the device with a damaged power cord or plug.

• Keep the power cord away from heated surfaces.

• Always unplug the device from the power outlet after use and before 

cleaning.

• Plug the device into the appropriate voltage outlet. Do not overload power 

outlets or use extension cords.

• Do not misuse the power cord.

• Do no wind the power cord around the compressor.

(Cleaning and disinfecting)

Observe the rules below when cleaning or disinfecting parts. Failure to 

observe these rules may result in damage, inefficient nebulization or 

infection. For the instructions, refer to “Cleaning and daily disinfecting” 

section.

• Clean and disinfect the nebulizer kit, baby mask, mouthpiece and nasal 

aspirator before using them when as follows:

- the first time after purchase

- if the device has not been used for a long period of time

- if more than one person uses the same device

• Be sure to wash or wipe the parts after use, and ensure that they are 

thoroughly disinfected and dried, and stored in a clean location.

• Do not leave the cleaning solution in the parts.

• Do not use a microwave oven, dish dryer or hair dryer to dry the device or 

the parts.

• Do not use an autoclave, EOG gas sterilizer or low temperature plasma 

sterilizer.

• When disinfecting parts by boiling, make sure that the container does not 

boil dry. Otherwise it may also result in fire.

 

Caution:

Indicates a potentially hazardous situation which if not avoided, 

may result in minor or moderate injury, or physical damage.

General Safety Precautions:

• Inspect the device and parts before using them each time, and check that 

there are no problems. In particular, be sure to check the following:

- That the nozzle of either the nebulizer kit or aspirator or air tube are not 

damaged or blocked.

- That the compressor operates normally.

• When using this device, there will be some noise and vibration caused 

by the device. This is normal and does not indicate a malfunction.

• Do not use if the device or any parts are damaged. Replace damaged 

parts.

• Do not subject the device or the parts to any strong shocks such as 

dropping the device on the floor.

• Do not insert objects inside the compressor.

• Do not disassembly or attempt to repair the compressor or power cord.

• Do not leave the device or its parts in extreme hot or cold temperature, 

or under direct sunlight. 

Keep these instructions for future reference.

Optional Medical Accessories

(within the scope of EC Medical Device Directive 93/42/EEC)

Product description

Model

DuoBaby Nebulizer accessory set

(contents: nebulizer kit (includes two inhalation tops), 

baby mask, mouthpiece, 3 air filters, air tube)

NEB2012

DuoBaby Aspirator accessory set, (contents: nasal 

aspirator, air tube)

NEB7000

Other Optional/Replacement Parts

Product description

Model

Air Filter set (contents: 3 pieces)

3AC408

Thank you for purchasing the Omron DuoBaby.

This product was developed in conjunction with respiratory therapists for the 

successful treatment of asthma, chronic bronchitis, allergies, blocked nose, 

and other respiratory disorders. This is a medical device. Operate the device 

only as instructed by your doctor and/or respiratory therapist.

How the nebulizer kit works

The medication that is pumped up through the 

medication channel is mixed with compressed 

air which is generated by a compressor pump. 

The compressed air mixed with medication is turned 

into fine particles and sprayed when in contact with 

the baffle.

How the nasal aspirator kit works

The compressed air generated by the 

compressor pump, passing through a 

narrowing, thanks to the Venturi effect, 

creates a vacuum which sucks the mucus 

into the collection chamber.

3A HEALTH CARE S.r.l.

Via Marziale Cerutti, 90F/G  

25017 Lonato del Garda (BS)  

Italy

Distributor

OMRON HEALTHCARE EUROPE B.V.

Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS

www.omron-healthcare.com

Subsidiaries

OMRON HEALTHCARE UK LTD.

Opal Drive, Fox Milne, Milton Keynes, MK15 0DG, UK

www.omron-healthcare.com

OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH

Gottlieb-Daimler-Strasse 10, 68165 Mannheim, GERMANY

www.omron-healthcare.com

OMRON SANTÉ FRANCE SAS

14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE

www.omron-healthcare.com

Made in Italy

Baffle

Aerosol

Medication 

Aerosol

Nozzle

Medication 

Compressed Air

Vacuum

Compressed Air

(Nebulizer: Usage)

• Provide close supervision when this device is used by, on, or near children, 

invalids or persons who cannot express their consent.

• Make sure that the parts are attached or installed correctly.

• Make sure that the air filter is clean and correctly attached. If the air filter 

has been used for more than 70 applications, replace it with a new one. 

• Do not tilt the nebulizer kit at an angle of greater than 60 degrees in all 

directions or shake it while in use.

• Do not use or store the device while the air tube is creased.

• Do not add more than 12 ml of medication to the medication tank.

• Do not carry or leave the nebulizer kit while the medication tank contains 

medication.

• This product should not be used on patients, who are unconscious, 

sleeping, are drowsy or are not breathing spontaneously.

• Do not distort the vaporiser head, or jab the nozzle of the medication tank 

with a pin or any sharp object.

• The compressor may become warm during operation, therefore do not 

touch it other than turning it off after nebulizing.

• To avoid medication residue on the face, be sure to clean the patient’s face 

after removing the mask.

(Nasal Aspirator: Usage) 

• The nasal aspirator should be used exclusively for removing the nasal 

mucus from a baby’s nose. It cannot be used for any other purpose.

• Do not obstruct the holes on the bottom of the nasal aspirator.

• If you notice rashes or redness appearing, inside or around the nose of the 

baby during or after using the device or anything unusual during use, stop 

using the device and consult a physician.

• Only use the original parts as included in the product.

• Do not attach the nasal aspirator to any other device.

DISPOSAL PROCEDURE (Dir. 2012/19/EU-WEEE)

This product is not to be treated as regular household waste but 

must be returned to a collection point for recycling electric and 

electronic devices. Further information is available from your 

municipality, your municipality’s waste disposal services, or the 

retailer where you purchased your product.

Compressor Nebulizer 

with Nasal Aspirator

DuoBaby (NE-C301-E)

Instruction Manual

IM-NE-C301-E-EN-01-05/2016

3A3427 rev.00

Rear View

Power Cord

Accessories 

Compartment 

with Lid

Power Plug

Air Filter Cover 

Including Air Filter

Ventilation Slots

Power Switch Air Tube 

Connector

Nebulizer Kit and 

Aspirator Holder

Compressor (Main Unit)

Baby Mask

Mouthpiece

Air Tube

Instruction Manual

Nebulizer Kit

Inhalation Top x 2

Upper Airways

Lower Airways

Vaporiser Head

Nozzle

Medication Tank

Air Tube Connector

Nasal Aspirator

1

2

Air Tube Connector

Finger 

opening

Nozzle

O-ring

Collection 

chamber