Removing condensation from the air tube

If there is moisture remaining in the air tube, be sure to follow the procedure 

below.

•  Make sure that the air tube is connected to the air connector on the compressor.

•  Remove the air tube from the nebulizer kit or the nasal aspirator.

•  Turn on the compressor and pump air through the air tube to expel the moisture.

Changing the air filter

If the air filter has changed colour, or has been used for more than 70 

applications, replace it with a new one. (see Other 

Optional / Replacement Parts).

1.

  Remove the air filter cover as shown in figure.

2.

  Remove the old air filter with a sharp object,  

such as a toothpick, and insert the new air filter.

3.

  Put the air filter cover back in place.

 

Notes:

•  Use only OMRON air filters designed for this device. Do not operate without a filter.

•  Do not attempt to wash or clean the filter. If the air filter becomes wet, replace it. 

Damp air filters can cause blockages and may increase risk of infection.

•  There is no front or back orientation for the air filters.

•  Check that air filters are clean and free of dust before inserting them.

1.

  Make sure that the power switch is in the off ( 

 ) position.

2.

  Assemble the nasal aspirator. Make sure that the O-ring is properly 

attached. Refer to the drawing “Nasal Aspirator”.

3.

  Plug the power plug into a power outlet.

4.

  Connect the air tube to the nasal aspirator and 

to the main unit, while twisting the air tube plugs 

slightly, push them firmly.

5.

  Keep your baby in a comfortable position and 

make sure that their head is lifted as shown on 

the figure to the right.

6.

  Turn the power switch to the on ( ) position. 

Place your index finger over the opening on the 

nasal aspirator, as shown on the figure to the right 

to begin aspiration.

 

Caution:

Before using the nasal aspirator, try it on the hand to 

see if it sucks. When functioning properly, there must be 

a slight suction from the nozzle. If air is exiting from the 

nozzle, do not continue using the aspirator.  

While using, the aspirator needs be in the upright position, with the finger opening 

on top (see figure by above). This prevents the sucked mucus from entering parts 

other than the collection chamber.

7.

  Place the nozzle at the entrance of one of the nostrils of the baby. (do not 

insert inside the nostril). Use the aspirator for 3 seconds for 4 times per 

nostril.

8.

  When treatment is completed, turn the power off and unplug the 

compressor from the power outlet.

9.

  Disconnect the aspirator from the air tube and disassemble it into 3 parts as 

shown in “Know your unit”; clean or disinfect according to the section below.

Cleaning and daily disinfecting
Cleaning

Clean the parts after each use to remove residual medication or mucus. 

This will prevent inefficient nebulization or aspiration and reduce the risk of 

infection. Do not use a brush or pin to clean.

•  Washable parts: 

Nebulizer kit, Baby Mask (PVC), Nasal Aspirator, Mouthpiece 

 

Wash them in warm water and mild, neutral detergent. Rinse them 

thoroughly with clean hot tap water, gently tap to remove excess water 

and allow to air dry in a clean place. It is advisable to replace the DuoBaby 

nebulizer kit or nasal aspirator after some 100 to 120 treatments or after 

about 20 boiling cycles.

• Non-washable parts: 

Compressor, Air Tube (PVC) 

Firstly, make sure that the power plug is unplugged from the power outlet. 

Wipe clean with a soft cloth, moistened with water or mild, neutral detergent. 

Air Filter 

Do not wash or clean the air filter. Refer to “Changing the air filter” section.

Disinfecting

Always disinfect the parts after the last treatment of the day. If the parts 

are heavily stained, replace them with new ones. To select a method for 

disinfection, refer to the table below.

o

: applicable 

X

: not applicable

Disinfecting 

method

Parts

Mouthpiece

Nasal  Aspirator

Nebulizer  Kit

Baby Mask

Air T

ube

Air Filter

Materials

PP

PP

PP

Mask: PVC  (Phthalate 

free)

PVC

(Phthalate 

free)

A

See  

notes 

below

Disinfecting ethanol

o

o

o

o

X

X

Sodium hypochlorite

o

o

o

o

X

X

Milton*

o

o

o

o

X

X

Quaternary ammonium

o

o

o

o

X

X

Osvan*

Chlorhexidine

o

o

o

o

X

X

Hibitane*

Amphoteric Surfactant

o

o

o

o

X

X

Tego*

B

Boiling

o

o

o

X

X

X

 

* an example of commercially available disinfectant.

Notes:

•  Use a commercially available disinfectant. Follow the instructions provided by the 

disinfectant manufacturer.

•  Never clean with benzene, thinner or a flammable chemical.

•  After having washed the Nebulizer Kit and/or Nasal Aspirator with the available 

disinfectant, please rinse with clean warm water and dry thoroughly before using.

•  Applicable parts may be boiled between 15 to 20 minutes.

•  After boiling, carefully remove the parts, shake off excess water and allow to air  

dry in a clean environment.

Troubleshooting for: main unit, nebulizer kit and nasal 

aspirator 

Check the following if the device should fail before or during operation.

Problem

Cause

Remedy

Nothing happens 

when the power 

switch is turned.

Is the power plug plugged into 

a power outlet?

Check that the plug is inserted in

a power outlet. Unplug then 

reinsert the plug if necessary.

No nebulization/

aspiration or low 

nebulization, when 

the power is on.

Is there sufficient medication 

in the medication tank? 

(only nebulizer kit)

Add the correct amount of 

medication to the medication tank.

Is the vaporizer head missing 

or not assembled correctly? 

(only nebulizer kit)

Attach the vaporizer head 

correctly.

Is the nebulizer kit or nasal 

aspirator assembled correctly?

Assemble the nebulizer kit or 

nasal aspirator correctly.

Is the nozzle for the nebulizer 

kit or aspirator blocked?

Make sure that the nozzle is free 

of blockages, do not pinch the 

nozzle.

Is the nebulizer kit or nasal 

aspirator tilted at a sharp 

angle?

Make sure that the nebulizer kit 

is tilted at an angle less than 

60 degrees, or the aspirator is 

upright.

Is the air tube connected 

correctly?

Make sure that the air tube is 

correctly connected to the

compressor and nebulizer kit or 

nasal aspirator.

Is the air tube folded, blocked 

or damaged?

Make sure that the air tube 

does not contain kinks or is not 

damaged or blocked.

Is the air filter dirty?

Replace the air filter with a new 

one.

The compressor is 

abnormally loud.

Is the air filter cover attached 

correctly?

Attach the air filter cover correctly.

The compressor is 

very hot.

The compressor is 

insufficiently ventilated.

Do not cover the compressor 

during use.
Do not block the ventilation slots.
Do not use the device in an 

environment with temperatures 

higher than 40°C. 

The compressor 

suddenly stops 

during use.

The thermal cut-out has 

shut the device down due to 

overheating

See note below.

Note:

Do not attempt to repair the device. Do not open and/or tamper with the device. 

No parts of the device are user serviceable. Return the device to an authorized 

OMRON retail outlet or distributor.

Technical data

Product Description:

Compressor Nebulizer with Nasal 

Aspirator

Model (code):

DuoBaby (NE-C301-E)

Rating:

230 V ~ 50 Hz

Power Consumption:

140 VA

Operating Temperature/ 

Humidity/Air Pressure:

+10°C to +40°C / 10% to 95%RH / 

690 - 1060 hPa

Storage and Transport 

Temperature/Humidity/

Air Pressure:

-20°C to +70°C / 5% to 95%RH / 

690 - 1060 hPa

Weight:

Approx. 1.7 Kg (compressor only)

Dimensions:

Approx. 170 (W) x 280 (D) x 105.5 (H) mm

Contents:

Compressor, Nebulizer Kit, Air tube, 

Mouthpiece, Baby Mask, Nasal Aspirator, 

Instruction Manual.

How to use nebulizer kit

1.

  Make sure that the power switch is in the off ( 

 ) 

position.

2.

  Plug the power plug into a power outlet.

3.

  Remove the inhalation top from the medication tank.

1) 

Rotate the inhalation top anti-clockwise

2) 

Lift the inhalation top out from the medication tank.

4.

  Add the correct amount of prescribed  

medication into the medication tank.

5.

  Verify the presence of the vaporiser  

head inside the medication tank.

6.

  Choose the appropriate inhalation top by reading the 

“User Guidance”.

  To put the inhalation top back onto the medication tank.

1) 

Lower the inhalation top onto the medication tank.

2) 

Rotate the inhalation top clockwise.

7.

  Attach the Baby mask 

or mouthpiece to the 

nebulizer kit tightly.

8.

  Attach the air tube. While twisting the air 

tube plug slightly, push it firmly into the air 

tube connector.

9.

  Hold the nebulizer kit as 

indicated on the right. Follow 

the instructions of your doctor 

or respiratory therapist.

 

Caution:

Do not tilt the nebulizer kit at an angle of greater than 60 degrees in all directions. 

Medication may flow into the mouth or it may result in ineffective nebulization.

10.

 Turn the power switch to the on ( ) position. As the compressor starts, 

nebulization begins and aerosol is generated. Inhale the medication. 

Exhale through the nebulizer kit.

11. 

When treatment is completed, turn the power off and unplug the 

compressor from the power outlet.

12. 

Clean and disinfect all (disassembled) parts according to the “Cleaning 

and daily disinfecting” section.

How to use Nasal Aspirator

Dry and very thick mucus inside the nose may be dissolved by saline 

solutions. After adding saline solution into each nostril of the baby’s nose 

to help loosen nasal secretions, wait a few minutes and then remove the 

mucus running out with a tissue. Only after removing the running mucus 

outside the nostrils the nasal aspirator may be used. 

1

2

Medication Tank

Vaporiser Head

1

2

Right angle

   =   Class  ll 

equipment

  =  Type BF 

applied part

  =  Power off

  Consult the 

instructions for use

IP 21

  =  Power on

~

  Alternating current

Nebulizer Kit

Appropriate Medication 

Quantities:

1 ml minimum - 12 ml maximum

Sound:

Noise level (at 1 m distance) 63 dB

Nebulizer  

Upper Airways

Nebulizer  

Lower Airways

Particle Size (MMAD*):

Approx. 10 μm

Approx. 3 μm

Nebulization Rate 

(by weight loss):

Approx. 0.4 ml/min

Approx. 0.2 ml/min

Aerosol Output 

(3 ml, 1% NaF):

Approx. 0.45 ml

Approx. 0.36 ml

Aerosol Output Rate 

(1% NaF):

Approx. 0.15 ml/min

Approx. 0.10 ml/min

*MMAD = Mass Median Aerodynamic Diameter

Result of cascade impactor measurements for particle size

Cumulative % particle mass of sodium fluoride undersize

Cumulative Undersize%

Particle Size Dp (μm)

Inhalation Top 1 – 

Green 

Color Upper Airways

Inhalation Top 2 – 

Blue 

Color Lower Airways 

Nasal Aspirator

Suction Pressure: 

Minimum -0.15 bar - Maximum -0.2 bar

Air Flow: 

Approx. 4 l/min

Maximum fill volume of  

collection chamber: 

7 ml

General Notes:

•  Subject to technical modification without prior notice.

•  The device may not work if the temperature and voltage conditions are different to 

those defined in the specifications.

•  The device fulfils the provisions of the EC directive 93/42/EEC (Medical Device 

Directive) and the European Standard EN13544-1:2007+A1:2009, Respiratory 

therapy equipment - Part1: Nebulizing systems and their components.

•  Performance may vary with drugs such as suspensions or high viscosity. See drug 

supplier’s data sheet for further details.

•  See website of OMRON HEALTHCARE EUROPE for complete up to date information. 

URL: www.omron-healthcare.com

Warranty

Thank you for buying an OMRON product. This product is constructed of high quality 

materials and great care has been taken in its manufacturing. It is designed to give 

you a high level of comfort, provided that it is properly operated and maintained as 

described in the instruction manual.

This product is guaranteed by OMRON for a period of 3 years after the date of 

purchase. The proper construction, workmanship and materials of this product is 

guaranteed by OMRON. During this period of guarantee OMRON will, without charge 

for labour or parts, repair or replace the defect product or any defective parts.

The guarantee does not cover any of the following:

a. Transport costs and risks of transport.

b. Costs for repairs and / or defects resulting from repairs done by unauthorised persons.

c.  Periodic check-ups and maintenance.

d. Failure or wear of optional parts or other attachments other than the main device 

itself, unless explicitly guaranteed above.

e. Costs arising due to non-acceptance of a claim (those will be charged for).

f.  Damages of any kind including personal caused accidentally or from misuse. 

Should guarantee service be required please apply to the dealer whom the product 

was purchased from or an authorised OMRON distributor. For the address refer to 

the product packaging / literature or to your specialised retailer.

If you have difficulties in finding OMRON customer services, contact us for information.

Repair or replacement under the guarantee does not give rise to any extension or 

renewal of the guarantee period.

The guarantee will be granted only if the complete product is returned together with the 

original invoice / cash ticket issued to the consumer by the retailer. OMRON reserves 

the right to refuse the guarantee service if any unclear information has been given.

User Guidance about the Nebulizer Kit

The DuoBaby nebulizer is a medical device that generates an aerosol with adjustable 

characteristics to adapt to the patient’s respiratory pathology

Inhalation Top 2 – Blue Color for 

Lower Airways: 

Treats conditions 

such as: asthma, bronchitis, 

bronchiolitis, bronchiectasis and 

bronchopneumonia.
Particle size (MMAD) is 

approximately 3 μm.

Inhalation Top 1 – Green Color for 

Upper Airways:

 Treats conditions 

such as: rhinitis, sinusitis, pharyngitis, 

tonsillitis, and laryngitis.
Particle size (MMAD) is 
approximately 10 μm.

Reviews: