42
ZM305 GTN Inbound-Outbound User Instructions
Supervisor menu
4.
Press
SELECT
…
The current setting is displayed (
oFF
or
on
).
5.
Press
UNITS
to toggle between the choices and when your choice is
displayed, press
ZERO
to accept …
Annun
is displayed.
Outputs
Setpoint
Edit
Annun
Out
6.
Press
UNITS
…
out1
is displayed. This is the weight value for setpoint 1. You can access
out2
or
out3
by pressing the
UNITS
key. The following instructions
apply to any of these three outputs.
7.
Press
SELECT
…
Edit X
is displayed.
X
being the number of output.
8.
Press
UNITS
…
ModE X
is displayed. This menu item sets the function of the output.
Mode selection must be made before entering the Out value.
9.
Press
SELECT
…
The current Mode setting for the selected Output is displayed. The Mode
settings are listed below:
Act AbV
(default)
The output is active when the weight is above the
set value
Act bEL
The output is active when the weight is below the
set value
Act in
The output is active when the weight is inside of
the low and high set values
Act out
The output is active when the weight is outside of
the low and high set values
10.
Press
UNITS
or
to scroll through the choices shown above and then
press
ZERO
to accept the displayed Mode …
ModE X
is displayed.
11.
Press
UNITS
…
Edit X
is displayed. Set the value or values for the output under this
menu item.
12.
Press
SELECT
…
If you selected Mode
Act AbV
or
Act bEL
, a value entry screen is
displayed. Go to step
13
.
If you selected Mode
Act in
or
Act out
,
outX Lo
is displayed. This is one
of two value entry screens under
Edit X
. Go to step
14
.