4
Calibration procedure
8
User Instructions for TB/TC Series Balances
4
Calibration procedure
You can set the accuracy of calibration either at full load or at
intermediate weight values. Use one of the following weights
during calibration:
TB6l
TB151
TC60
200g
500g
2000g
500g
1000g
5000g
600g
1500g
6000g
The weight used should ideally correspond with the operable
weight. If the balance is to be used for general weighing over a
wide range, select the maximum weight.
4.1
Calibrating the balance
1.
Ensure that the pan is empty and the display shows zero.
2.
Place the required calibration weight centrally on the pan.
3.
Press and hold
g
g
. Press
T
until the display
shows CAL and then release both keys.
The display shows CAL for a short while and then displays
the value of the calibration weight used. This signifies that
calibration is complete.
4.
Wait until calibration is complete and then remove the
weight from the pan.
5.
Check that the balance is calibrated accurately by ensuring
that the display shows
D
zero when no weight is on the pan
D
the correct value when a calibration weight is applied
6.
If the balance is not accurately calibrated repeat from
step 1.
Summary of Contents for TB Series
Page 1: ...76101 241 Issue 5 21 07 97 Electronic Top Pan Balances TB TC Series User Instructions ...
Page 2: ......
Page 23: ......