5
Setup procedure
12
User Instructions for TB/TC Series Balances
3.
Press
g
g
to display the required option.
4.
To return to normal weighing press
T
.
5.
To configure data send see section 5.1.1.
Configuring kilogrammes/hectolitre
If H–Itr is selected as the secondary weighing unit (see
section 5.2) the balance calculates kilogrammes per hectolitre.
1.
Ensure you are in setup mode.
2.
Press
g
g
to display H–Itr.
3.
Place a 500ml volumetric container on the pan and zero the
balance.
4.
Fill the container with the grain under test, using a
repeatable filling method.
5.
Replace the container and contents on the pan.
The display shows the sample weight compared with its
volume expressed as kilogrammes/hectolitre.
Summary of Contents for TB Series
Page 1: ...76101 241 Issue 5 21 07 97 Electronic Top Pan Balances TB TC Series User Instructions ...
Page 2: ......