Verifying System Operation
Use the following procedure to verify that the alarm systems
are working properly.
1. Set up Aulisa GA2000 system according to instructions above. (See
“Setting up Aulisa GA2000 system” section on page 21 for more
information.)
2. Ensure there is a system connection established between the Oximeter
Box, Receiver/Transponder and Display Unit. (See “System Connection”
section on page 25 for more information.)
3. Detach the sensor cable from Oximeter Box.
4. Verify that an alarm message is displayed and that an alarm audio is
generated (See “Troubleshooting” section on page 49 if an alarm
message and audio signal is not generated.)
5. Press on the “PAUSE AUDIO” button to temporarily silence for 2
minutes.
6. After alarm signal is regenerated, press on the “AUDIO OFF” button to
silence permanently the alarm signal.
NOTE: Alarm systems should be checked before use.
Use the procedure below to monitor the readings (SpO
2
, pulse rate,
pulse amplitude) in order to verify that the device is functioning properly.
1. Set up Aulisa GA2000 system according to instructions above (See
“Setting up Aulisa GA2000 system” section on page 21 for more
information.)
2. Press firmly on the application site to make sure it securely attached to
the finger.
3. Verify the Bluetooth connection status icon on the Display Unit is blue
and the power On/Off button on the Oximeter Box is blinking green.
30
Summary of Contents for Guardian Angel Rx GA2000
Page 1: ...Guardian Angel Rx GA2000 Digital Vital Sign Monitoring System Instructions For Use TM...
Page 11: ...Adult Oximeter Sensor Cable Wristband Pediatric Oximeter Sensor Cable 8...
Page 12: ...Stand Display Unit Charging Adaptor Oximeter Box Charging Adaptor Display Unit 9...
Page 39: ...O X I M E T E R B O X B AT T E RY L O W John 36...
Page 66: ...Service Support and Warranty Warranty under Contact Us 63...
Page 68: ...65...
Page 72: ...information aulisa com 69...
Page 79: ...7MN00005 01...