31
Medium priority audio alarms are: 3 beeps and 25-second pause. This sequence
repeats until the alarm is cleared or silenced.
Tap on “PAUSE AUDIO” button to pause the alarm audio for 2 minutes. Tap on
“AUDIO OFF” button to permanently turn off the alarm audio.
See “Alarm Limits” section to learn adjusting the alarm limit.
NOTE:
The following table describes alarm conditions and visual indicators.
Alarm Condition
(Medium Priority Alarm)
Visual Indicator
Oximeter Sensor Probe Detached from the
Finger
The Oximeter Module blinks yellow
along with blinking yellow displays
and alarm text message on the
Display Unit.
Oximeter Module Battery Low
Oximeter Sensor Cable Detached
Oximeter Module Detection Error
Display Unit Battery Low
On the Display Unit, it blinks yellow
displays and alarm text message.
Thermometer Battery Low
Oximeter Module Data Update Period Exceeds
Limit (More Than Half A Minute)
Aulisa sensor module(s) BLE Disconnected
Receiver/Transponder Wi-Fi Disconnected