7. Test Descriptions
7.4.2
Voltage Tests
RED 70Vrms Sine Wave Recognition
Test Description :
Verify that a sine wave greater than 70Vrms is recognized as being "on" for every RED channel input.
Test Sequence :
1. Monitor is powered up in the quiescent state.
2. Redenable input is activated.
3. A sine wave greater than 70Vrms is applied to each RED channel.
Monitor Response to PASS
:
Monitor must NOT be found in fault state.
RED 70Vrms Positive Rectified Recognition
Test Description :
Verify that a positive rectified sine wave greater than 70Vrms is recognized as being "on" for every RED channel input.
Test Sequence :
1. Monitor is powered up in the quiescent state.
2. Redenable input is activated.
3. A positive rectified sine wave greater than 70Vrms is applied to each RED channel.
Monitor Response to PASS
:
Monitor must NOT be found in fault state.
RED 70Vrms Negative Rectified Recognition
Test Description :
Verify that a negative rectified sine wave greater than 70Vrms is recognized as being "on" for every RED channel
input.
Test Sequence :
1. Monitor is powered up in the quiescent state.
2. Redenable input is activated.
3. A negative rectified sine wave greater than 70Vrms is applied to each RED channel.
Monitor Response to PASS
:
Monitor must NOT be found in fault state.
RED 50Vrms Sine Wave Recognition
Test Description :
Verify that a sign wave less than 50Vrms is recognized as being "off" for every RED channel input.
Test Sequence :
1. Monitor is powered up in the quiescent state.
2. All RED channels are at AC line voltage.
3. Redenable input is activated.
4. A sine wave less than 50Vrms is applied to each RED channel one at a time.
5. fault state is checked for each channel.
Monitor Response to PASS
:
PCMT-8000 User's Manual