Chill V Series

5. Operating Panel (con’t)

Set Value Setting:

In basal display status, we can set the set-point (SV) by pressing   

 

        or            .

Press    

 to decrease the value,             to increase the value, and                to move  

to the digit expected to modify. Keep pressing             or            ,  to speed up the decreasing  
or increasing of the value. 

Parameter Setting:

In basal display status, press            and hold for about two (2) seconds in order  
to access Field Parameter Table. Pressing              can go to the next parameter;  
pressing                          or               can modify a parameter. The instrument will escape 
automatically from the parameter table if no key is pressed within 30 seconds.

Intelligence control and auto tuning

When the “At” control method is chosen, the control parameters can be obtained by  

running auto-tuning. At the first time of running auto-tuning, in basic display status, “At” 

 

will flash in the lower display window and the instrument executes on-off control. After 

 

two (2) to three (3) cycles of on-off action, the instrument will obtain the values of PID  
control parameters. If you want to escape from auto tuning status, press              for about  

two (2) seconds until the “At” flash stops. Depending on the system, the time of auto tuning 

 

can be several seconds to several hours. After the auto tuning finishes, the parameter “At” 

automatically changes to “OFF”. If the user wants to run auto tuning again, set the setting 
parameter  from “At” to “on”. 

Note 1: 

Before running auto-tuning, the set-point should be set to an often-used value  

or middle value first, and then start auto-tuning. Generally, for applications with heating 

 

time of 10~80% satisfying auto-tuning results can be obtained. If the heating time is not 

 

in the range of 10~80%, the auto-tuning will continually run, searching for satisfying 

 

result. Raising the set-point for heating time less than 10% or decreasing the set-point 

 

for heating time greater than 80% can improve auto-tuning result.

Note 2:

 Don’t modify the set-point during auto-tuning. It may affect the precision of auto-tuning.

Note 3:

 The temperature fluctuation caused by interference may affect the precision of 

 

auto-tuning. Besides checking the wiring, increasing the parameter FILt (input digital filter) 

 

can also reduce interference.

Note 4: 

After auto-tuning finishes, the temperature rising typically has a little over-shoot 

 

(about 1~3°C). The transducer is often placed between the heater and heated material,  
and a little over-shoot can shorten the time of rising to set-point and save power. (Generally,  
the temperature of the transducer can be raised quickerthan that of the heated material.)

Summary of Contents for Chill V Series

Page 1: ...ATS Chill V Series Re CirculatingChiller Instructional Manual 89 27 Access Road Norwood MA 02062 T 781 769 2800 F 781 769 9979 www qats com ATS Chill V Series...

Page 2: ...including but not limited to implied warranties of merchant ability and fitness for a particular purpose with regard to the Software and accompanying written and Hardware material This limited warrant...

Page 3: ...Contents 1 Safety Notes 3 2 Unpacking 3 3 Equipment Description 3 4 Brief Instructions 4 5 Operating Panel 5 6 Technical Data 8 7 Maintenance 8 Table of Contents...

Page 4: ...or servicing or repair before moving the equipment Drain the bath before moving the equipment Have the equipment serviced or repaired only by properly qualified personnel Other ATS Liquid Cooling Prod...

Page 5: ...cooled through a heat exchanger located inside the bath Heat of condensation and motor heat are dissipated on air cooled units through a fan cooled ribbed condenser The refrigeration unit is protecte...

Page 6: ...the pressure connection should be left open when the equipment is filled for the first time otherwise there may be permanent damage to the pump Remove the plug on the filler opening at the upside A f...

Page 7: ...contact Check the details on the rating label against the supply voltage No warranty applies when the equipment is connected to the wrong supply 4 4 Emptying Place a container underneath the drain co...

Page 8: ...Lamp AUTO Auto Optimize Auto tuning OVER Over temperature alarm LEVEL Low Level alarm Option Operation Description Basal display status When powered on the upper display window of the instrument shows...

Page 9: ...e several seconds to several hours After the auto tuning finishes the parameter At automatically changes to OFF If the user wants to run auto tuning again set the setting parameter from At to on Note...

Page 10: ...p Flow Rate 7 10 L min 15 L min 10L min 10 L min Working mode Continuous Continuous Continuous Continuous Filling Port mm 25 25 25 25 Reservoir Capacity 1 0 L 4 5 L 8 L 8 L Dimensions W x D x H 230 x...

Page 11: ...here the condenser of the refrigerator has to be cleaned at intervals of 4 to 6 months or more frequently This is best done by blowing compressed air or nitrogen into the ventilation openings for a fe...

Page 12: ...ality test equipment s and leading edge R D specifically tailored to the telecom LED and computing industries In addition ATS provides thermal design consulting services and training for the electroni...

Reviews: