Figure 1 Water Outlets
Figure 2 Water Tank
Container
Section 2: Unpacking
The ATS-Chill V Series chillers are packed carefully to prevent
damage in transport. After unpacking, first check the equipment and
accessories for possible transport damage. If, unexpectedly, the
equipment is noticeably damaged, it is essential to notify the forwarding
agent or the postal authorities so an inspection can take place.
Do not
start operations with damaged instruments.
The unit may NEVER
be overturned nor put upside down
Section 3: Equipment description
Materials
All components in contact with the thermostating liquid are made
specifically from materials suitable for the recommended liquids and
temperature, such as rust free, stainless steel, copper, brass, NBR
and/or plastics.
Refrigeration unit
The refrigeration unit consists essentially of a fully sealed
compressor. The thermostating liquid is cooled through a heat
exchanger located inside the bath. Heat of condensation and motor
heat are dissipated on air-cooled units through a fan-cooled ribbed
condenser.
The refrigeration unit is protected against overpressure and
compressor overload. With insufficient ventilation of the condenser
(e.g. dirt deposits) the equipment is switched off.
Control unit
The control unit is provided with an LED display which indicates
the measurements and settings as well as the operational status
functionality. The set-point and other control settings are selected
using two or three keys.
A Pt 100 temperature probe senses the outflow temperature inside
the bath. The measurement is processed by a high-resolution AD
converter. Further processing involves a special control algorithm to
operate the heating triac (with reduced reaction on the main supply),
the automatic compressor control, and the cooling control which
operates with low-noise solenoid valves. The proportional cooling
principle permits control in the cooling range without any energy
wasting counter-heat.
Chill V Series