5. Operating Panel
Key Functions
POWER
Power key
COOL
Cooling Key
PUMP
Circulating Key
Setup key, for accessing parameter
table and confirming change
Data shift key, also for
activating auto turning
Date increase key
Data decrease key
Display window
PV
Process Value Display: Displays PV or code of a parameter
SV
Setting Value Display: Displays SV, alarming code or value of a parameter
LED indicator
OUT
Output Lamp
AUTO
Auto Optimize (Auto-tuning )
OVER
Over temperature alarm
LEVEL
Low Level alarm (Option )
Operation Description
Basal display status:
When powered on, the upper display window of the instrument
shows the process value (PV), and the lower window shows the set-point (SV). This status
is called basal display status. When the input signal is out of the measurable range (for
example, the RTD circuit is broken, or input specification is set wrong), the upper display
window will alternately display “orAL” and the high limit or the low limit of PV, and the
instrument will automatically stop output.
Chill V Series