6. Technical Data

Specifications for Chill-V

Model

Chill150V 

Chill300V 

Chill600V

Chill1000V

Temperature range 

@25°C

5~40°C

5~35°C

5~35°C

5~35°C

Temperature Stability

±0.1°C 

±0.1°C 

±0.1°C

±0.1°C

Cooling 

Capacity@20°C     

150W

300W

600W

1000W

Pump pressure        

0.4bar

0.6/4bar

4bar

4bar

Pump Flow Rate   

7-10 L/min

15 L/min

10L/min

10 L/min

Working  mode              

Continuous

Continuous

Continuous

Continuous

Filling Port  mm        

Ø25

Ø25

Ø25

Ø25

Reservoir Capacity

1.0 L

4.5 L

8 L

8 L

Dimensions   

(W x D x H)          

230 x 260 x 380mm

(9.06 x 10.24 x 14.96”)

235 × 475 × 490mm 

(9.25 x 10.24 x 19.29")

325 x 620 × 550 

(12.8 x 24.4 x 21.65")

325 x 620 × 550 

(12.8 x 24.4 x 21.65”)

Power  Volts  Hz  Amps              AC220V, 50Hz, 5A AC220V, 50Hz, 10A AC220V, 50Hz, 10A AC220V, 50Hz, 10A

Shipping Weight  (kg)

10

23

32

40

Chill V Series

Summary of Contents for Chill V Series

Page 1: ...ATS Chill V Series Re CirculatingChiller Instructional Manual 89 27 Access Road Norwood MA 02062 T 781 769 2800 F 781 769 9979 www qats com ATS Chill V Series...

Page 2: ...including but not limited to implied warranties of merchant ability and fitness for a particular purpose with regard to the Software and accompanying written and Hardware material This limited warrant...

Page 3: ...Contents 1 Safety Notes 3 2 Unpacking 3 3 Equipment Description 3 4 Brief Instructions 4 5 Operating Panel 5 6 Technical Data 8 7 Maintenance 8 Table of Contents...

Page 4: ...or servicing or repair before moving the equipment Drain the bath before moving the equipment Have the equipment serviced or repaired only by properly qualified personnel Other ATS Liquid Cooling Prod...

Page 5: ...cooled through a heat exchanger located inside the bath Heat of condensation and motor heat are dissipated on air cooled units through a fan cooled ribbed condenser The refrigeration unit is protecte...

Page 6: ...the pressure connection should be left open when the equipment is filled for the first time otherwise there may be permanent damage to the pump Remove the plug on the filler opening at the upside A f...

Page 7: ...contact Check the details on the rating label against the supply voltage No warranty applies when the equipment is connected to the wrong supply 4 4 Emptying Place a container underneath the drain co...

Page 8: ...Lamp AUTO Auto Optimize Auto tuning OVER Over temperature alarm LEVEL Low Level alarm Option Operation Description Basal display status When powered on the upper display window of the instrument shows...

Page 9: ...e several seconds to several hours After the auto tuning finishes the parameter At automatically changes to OFF If the user wants to run auto tuning again set the setting parameter from At to on Note...

Page 10: ...p Flow Rate 7 10 L min 15 L min 10L min 10 L min Working mode Continuous Continuous Continuous Continuous Filling Port mm 25 25 25 25 Reservoir Capacity 1 0 L 4 5 L 8 L 8 L Dimensions W x D x H 230 x...

Page 11: ...here the condenser of the refrigerator has to be cleaned at intervals of 4 to 6 months or more frequently This is best done by blowing compressed air or nitrogen into the ventilation openings for a fe...

Page 12: ...ality test equipment s and leading edge R D specifically tailored to the telecom LED and computing industries In addition ATS provides thermal design consulting services and training for the electroni...

Reviews: