USER’S MANUAL
Page 30 of 39
P000463.F
•
Storage temperature:
–35
C to +54
C
•
Humidity:
15 to 90% relative humidity
7.4.
Electrical Specifications
•
100-120V
~
50/60 Hz
•
220-240V
~
50/60 Hz
7.5.
Fuses
•
100 -120V, 220-240V, ~50 / 60 Hz,:
Replace fuses as marked:
1.25A/250V, T-lag, 5
20 mm, UL Recognized, IEC Approved
7.6.
Footswitch Specifications
•
Moisture protection rating:
IPX8
7.7.
Power and Voltage Output Restrictions
The maximum power output of 28.5 W for the Isolator
clamp is available at 114
load for
devices operating under device code “A, F, G, and H”. Lower maximum output powers are
available depending on the system operating mode. See Section 7.1.
The maximum power output of 15.0 W for the Isolator
Transpolar
pen is available
between 40
to 400
load for device operating under device code “B”. Lower maximum
output powers are available depending on the system operating mode. See Section 7.1.
The maximum power output of 30.0 W for the Coolrail™ linear pen is available between 47
to 200
load for device operating under device code “L”. Lower maximum output powers are
available depending on the system operating mode. See Section 7.1.
The maximum power output of 20.0 W for the Isolator™ linear pen is available between 31
to 300
load for device operating under device code “C”. Lower maximum output powers are
available depending on the system operating mode. See Section 7.1.
At other load impedances, the ASU will reduce the available power to comply with the
specified voltage and current limits. See Figure 11 and Figure 12.
The ASU is capable of producing a maximum output power of 32.5 Watts under a 100 Ohm
load although no current AtriCure
Bipolar Handpiece uses power above 30 Watts.
The maximum output voltage depends on the device code, and can be either 57 Vrms or
77.5 Vrms. See Section 7.1.
7.8.
Equipment Type / Classification
•
Class 1 Equipment