P000463.F
Page 29 of 39
7.
Technical specifications
7.1.
RF Output
•
Frequency:
460 kHz
5%, Quasi-sinusoidal
•
ASU Maximum Power Output:
32.5 W at 100
•
HF Power and Voltage Output:
Device
Code
Maximum
Output Power
Maximum
Output Voltage
Handpiece Type
A
28.5 W at 114
57.0 Vrms
Isolator
clamp
B
15.0 W from
20
to 400
77.5 Vrms
Isolator
Transpolar
pen
C
20.0 W from
31
to 300
77.5 Vrms
Isolator
Transpolar
pen
Isolator
linear pen
D
25.6 W at 127
57.0 Vrms
Isolator
clamp
E
22.8 W at 143
57.0 Vrms
Isolator
clamp
F
28.5 W at 114
57.0 Vrms
Isolator
clamp
G
28.5 W at 114
57.0 Vrms
Isolator
clamp
H
28.5 W at 114
57.0 Vrms
Isolator
clamp
J
12.0 W from
20
to 500
77.5 Vrms
Isolator
Transpolar
pen
K
25.0 W from
39
to 240
77.5 Vrms
Isolator
Transpolar
pen, or
Coolrail™ linear pen
L
30.0 W from
47
to 200
77.5 Vrms
Isolator
Transpolar
pen, or
Coolrail™ linear pen
7.2.
Mechanical Specifications
•
Size:
13
13.75
6
(32.5 cm
34.4 cm
15 cm) maximum.
•
Weight:
15 lb. (9 kg) maximum.
7.3.
Environmental Specifications
•
Operational temperature:
10
o
C to 40
o
C