USER’S MANUAL
Page 20 of 39
P000463.F
4.2.
Operating Modes
The ASU operates in one of five modes: STANDBY, READY, RF ON, ERROR and FAULT
modes. These modes are shown on the lower left corner of the Conductance Display Graph. See
Figure 6, below.
•
STANDBY Mode
– This mode is entered automatically after the ASU is successfully turned on or
from READY Mode upon detection of a Handpiece or Footswitch disconnection. The LCD
display message indicates the system is in the STANDBY Mode.
•
READY Mode
– This Mode is entered upon connecting both Handpiece and Footswitch while in
the STANDBY Mode or from the ON Mode if the Footswitch has been depressed and released.
The LCD display message indicates the system is in the READY Mode.
•
RF ON Mode
– This Mode is entered when the Footswitch is depressed while in the READY
Mode. The system transitions from the RF ON Mode to the READY Mode upon 40-second time
expiration or if the Footswitch is released.
•
ERROR
Mode
– This Mode is entered upon detection of any recoverable error conditions during
any Mode excluding the FAULT Mode. The system displays the corresponding error message,
and upon Footswitch release, transitions to the READY Mode.
•
FAULT Mode
– This Mode is entered upon detection of any unrecoverable error condition during
any Mode. The system is inoperable in this Mode until the power is cycled off, then on.
Figure 6 – Conductance and Power Display Graph Indicating STANDBY Mode